GE CARESCAPE - Calibrating the Dual CO2 sensor; Calibrating the Dual CO2 adapter

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CO2withCAPNOSTATMainstream,CapnoFlexLF,andDualCO2modules
9.Checkthemonitorscreen.IfthemessageCalibratesensorappears,calibratethe
sensor.IfthemessageCheck/calibr.adapterappears,calibratetheadapter.
10.Blocktheopenendofthesamplelineornasalcannula.Within90secondsthe
pumpwillshutoffandthemessageSamplelineblockedshouldappearonthe
monitordisplaytoensurethatthemonitorrecognizesblockages.
11.Toturnthepumpbackonandreturntonormaloperation,turnthepumpoffand
thenbackonagainintheCO
2
setupmenu.Thepumpwillturnbackonandthe
modulewillreturntonormaloperation.
12.Connectthesamplelinetothepatient.Ensurethatthetubingisnothanging
lowerthanthepatient.
NOTE
Deliveryofbronchodilatorsand/ormucolyticsviaaerosoland
meterdoseinhalerscanbecorrosiveandcausepremature
blockageofthewatertrap.Discontinuesidestreamgas
analysispriortotreatmentbyswitchingthepumpoffand
removingthesamplelinefromthebreathingcircuit.
CalibratingtheDualCO2sensor
SensorcalibrationshouldbeperformedwheneverthemessageCalibratesensor
appears.
1.PlacethesensorawayfromallsourcesofCO
2
,includingthepatient’sexhaled
breath,yourexhaledbreath,andventilatorexhaustvalves.
2.Withthecableconnectedtothemodule,selectagasrelatedparameterwindow>
Calibration.
3.Placethesensoronthecellmarked0andselectStartCalibrating.
4.AfterthemessageZeroedappears,removethesensorfromthe0cellandattach
ittothecellmarkedREF.
ThemeasuredEtCO
2
valueappearsintheConnectsensortoREFcellwindow.If
thevaluestaysbetween36and40mmHgforthreesecondsduringthe20-second
calibration,thecalibrationissuccessfulandthemessageCalibratedappears.Ifthe
calibrationfails,themessageRecalibrateappearsinstead.Inthiscase,perform
anewcalibration.
CalibratingtheDualCO2adapter
AdaptercalibrationshouldbeperformedwheneverthemessageCheck/calibr.
adapterappears.
1.PlacethesensorandadapterawayfromallsourcesofCO
2
,includingthepatient’s
exhaledbreath,yourexhaledbreath,andventilatorexhaustvalves.
2.Withtheadapterconnectedtothesensorandthesensorconnectedtothe
monitor(and,whenused,allsidestreamtubingattachedandthepumpturned
on),selectagasrelatedparameterwindow.
3.SelecttheCalibrationtab.
4.SelectStartCalibrating.
ThemessageCalibratedappearsonthescreenafterasuccessfulcalibration.Ifthe
calibrationfails,themessageCalibrationerrorappearsinstead.Inthiscase,perform
anewcalibration.
314CARESCAPEModularMonitors2062971-001