Messages
Message
Location
ExplanationWhattodo
●IsoelectricEEG
●param.,
wavef.
Isoelectric(atline)EEGdetected
inEntropymeasurement.
●Checkthepatientstatus.The
patient’sanestheticstatus
maybeunnecessarilydeep.
●Lowsignal
●param.
MeasuredEEGsignalistoolow
forreliableEntropycalculation.
●Checksensorplacement.
●Thepatientmaybeintotal
suppression;checkthe
patientstatus.
●NoEntropysensor
●al.areaThesensorisnotconnectedto
thecable,orthesensorand
cablearenotcompatible.
●Checkconnectionbetween
Entropysensorandcable.
●Checkthatthecableand
sensorarecompatible.
●Nosensor
●al.areaThesensorisnotconnectedto
thecable,orthesensorand
cablearenotcompatible.
●Checkconnectionbetween
Entropysensorandcable.
●Checkthatthecableand
sensorarecompatible.
●Noise
●wavef.UnreliableEntropycalculation
ordistortedEEGwaveformmay
appearduringelectrosurgeryor
otherhighfrequencynoise.
●InterpretEntropyvalueswith
caution.
●Sensorcheckfailed
●param.
Thesensorhasnotpassedthe
impedancecheck.
●Checksensorplacementand
attachment.
●Conrmpropercontactof
eachelectrodeinthesensor.
●Replacethesensor.
●Sensoroff
●param.
Thesensorisconnectedtothe
cablebutnotattachedtothe
patient,orthesensortypeisnot
correct.
●Checkthatthesensoris
properlyattachedtothe
patient.
●Checkthatthesensorisan
Entropysensor.
●Startingup
●param.
Themonitoriscollectingdatato
startthemeasurement.
●Waitforaboutoneminute.
Entropyvaluesappear
automatically.
MessagesrelatedtoNMTmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
498CARESCAPEModularMonitors2062971-001