Messages
●report=reportview
●wavef.=waveformarea
Message
Location
ExplanationWhattodo
●Artifacts
●param.,
wavef.
Signalscontainnoiseorartifact.●Checksensorcontact.
●Removesourcesofexcessive
noise.
●Automaticcheckoff
●param.
Automaticsensorcheckhas
beenturnedoff.
●Ifrequired,activatethe
automaticcheck.
●Cableoff
●param.
Entropycableisoff.●Connectthecable.
●Checkingsensor
●param.,
wavef.
Sensorcheckisinprogress.●Waituntilthecheckisover.
Checkresultsaredisplayed.
●Conrmelectrode1
●Conrmelectrode2
●Conrmelectrode3
●param.
Oneofthesensorelectrodeshas
poorcontact.
●Conrmpropersensor
electrode1,2or3contactas
indicatedinthemessage.
●Conrmelectrodes
●param.
Morethanoneofthesensor
electrodeshavepoorcontact.
●Conrmproperelectrode
contact.
●Entropycableoff●al.areaTheEntropysensorcableis
notconnectedtotheEntropy
module.
●ConnecttheEntropycableto
theEntropymodule.
●Entropy
measurement
removed
●al.areaAcquisitionmodulehasbeen
removed.
●Connectthemoduleif
youwanttorestartthe
measurement.
●EntropyREhigh
/EntropyRElow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Adjustdrugtitration.
●EntropySEhigh
/EntropySElow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Adjustdrugtitration.
●Entropysensorcheck
failed
●al.areaThesensorhasnotpassedthe
impedancecheck.
●Checksensorplacementand
attachment.
●Conrmpropercontactof
eachelectrodeinthesensor.
●Replacethesensor.
●Entropysensoroff●al.areaThesensorisconnectedtothe
cablebutnotattachedtothe
patient,orthesensortypeisnot
correct.
●Checkthatthesensoris
properlyattachedtothe
patient.
●Checkthatthesensorisan
Entropysensor.
●IdenticalEntropy
modules
●al.areaTherearetwoormoreEntropy
modulesinthesystem.
●RemoveallbutoneEntropy
module.
2062971-001CARESCAPEModularMonitors497