GE CARESCAPE - 12 Invasive pressures; Invasive pressures compatibility limitations; Invasive pressure safety precautions; Invasive pressure warnings

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Invasivepressures
Invasivepressurescompatibilitylimitations
TRAMandTram-RacmoduleswiththeB850only.
Fordetailedinformationregardingmodule,monitor,andaccessorycompatibility,see
theCARESCAPEModularMonitorsSupplementalInformationManual.
Invasivepressuresafetyprecautions
Invasivepressurewarnings
WARNING
Allinvasiveproceduresinvolveriskstothepatient.Useaseptic
technique.Incorrectuseofthecathetercanleadtovessel
perforation.Followcathetermanufacturer'sinstructions.
WARNING
Makesurethatnopartofthepatientconnectionstouches
anyelectricallyconductivematerialincludingearth.
WARNING
Mechanicalshocktoaninvasivebloodpressuretransducer
maycausesevereshiftsinthezerobalanceandcalibration,
andcauseerroneousreadings.
WARNING
Repositioningthepatientafteracompletedzeroingprocedure
maycauseincorrectmeasurementvalues.
WARNING
WhenconnectingPDMorTRAM,theloadedIPlabelsmay
affectthechannellabelingofotheralreadyconnected
channels,andconsequentlyalsothealarmlimits.
Invasivepressuremeasurementlimitations
E-modulesusedforthismeasurementarenotsuitableforusewithneonatal
patients.
Invasivepressurepointstonote
Dependingontheinvasivepressuremoduleused,notallinvasivepressure
measurementsandsettingsareavailabletovieworchange.