Calculations
SelectingthePCWPsource
1.SelectData&Pages.
2.SelectCalculations.
3.SelecttheHemotab.
4.SelectEditInput.
5.SelectthesourcefromthePCWPSourcelist:
PCWP
LAPmean
PAdia
Indexingparametersforhemodynamicand
oxygenationcalculations
Indexedvaluesarecalculatedonlyifthepatient’sBSA(bodysurfacearea)valueis
availableatthetimewhenthecalculationstakeplace.
1.SelectData&Pages.
2.SelectHemoorOxytab.
3.SelectView.
4.SelecttheIndexedcheckboxatthelowerpartoftheview.
Thoseparametersthatcanbeindexedarenowdisplayedasindexed,andindexed
valuesarecalculated.
Editingcalculationinputvalues
VentilationcalculationswiththeB850andB650only.
1.SelectData&Pages.
2.SelectCalculations.
3.SelectHemo,Oxy,orVenttab.
4.SelectEditInput.
5.EnteroredittheparametervalueswiththearrowsoftheValuecolumn.
6.Toperformtheactualcalculationandsavethevalues,selecttheViewtab>Save.
IfyouselectPreviousMenubeforesavingthevalues,theyarelost.
2062971-001CARESCAPEModularMonitors407