Messages
Message
Location
ExplanationWhattodo
●Zeroing
●param.
IPchanneliscurrentlybeing
zeroed.
●Noactionrequired.
Messageisautomatically
removedandreplacedwiththe
zeroingresultsaftercompletion.
●ZeroICPseparately●al.areaTheICPchannelmustbe
zeroedseparatelyfromallother
invasivepressures.
●Zerothechannelusingthe
Zerooptionfoundunderthe
ICPchannelsetupmenu.
Messagesrelatedtotemperaturemeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
Message
Location
ExplanationWhattodo
●Calibrationcheck
●param.
Thereisa0.1°Cdifference
betweenthemeasured
temperatureandtheinternal
calibrationfortheindicated
channel.
●Changethecable.
●Changethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Calibrationfail
●param.
Calibrationontheindicated
channelfailed.
●Checkconnections.
●Replacethetransducer.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Identicaltemperature
modules
●al.areaTherearetwoormoreidentical
temperaturemodulesinthe
system.
●Removeallbutone
temperaturemodule.
●Nosensordetected
●param.
Nosensordetected.●Checkthesensorand
connections.
●Performingtemptest
●param.
Moduleiscalibrating.●Noactionrequired.
●T1Calibrationcheck/
T2Calibrationcheck
●al.areaThereisa0.1°Cdifference
betweenthemeasured
temperatureandtheinternal
calibrationfortheindicated
channel.
●Changethecable.
●Changethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
2062971-001CARESCAPEModularMonitors481

Table of Contents