Messages
Message
Location
ExplanationWhattodo
●RVPsyshigh/RVPsys
low
●RVPmeanhigh/
RVPmeanlow
●RVPdiahigh/RVPdia
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●SensorP1failedto
SensorP8failed
●al.areaIP:Faultyordisconnected
sensor.
●Checkthecableand
connections.
●Replacethetransducer.
●Sensor
●param.
IP:Faultyordisconnected
sensor.
●Checkthecableand
connections.
●Replacethetransducer.
●UACdisconnect●al.areaInvasivepressurelineis
disconnected.
●Checkthepatientstatus.
●Checkconnections.
●Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
●UACsyshigh/
UACsyslow
●UACmeanhigh/
UACmeanlow
●UACdiahigh/UACdia
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●UVCsyshigh/
UVCsyslow
●UVCmeanhigh/
UVCmeanlow
●UVCdiahigh/UVCdia
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Zeroadj>100mmHg
●param.
IPchannelzeroedtoover100
mmHgpressure.
●Repeatthetransducer
zeroing.
●Replacethesensor.
●Replacethetransducer.
●Replacethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Zeroed
●param.
Zeroingwassuccessful.●Noactionrequired.
Messageisautomatically
removedafter10seconds.
480CARESCAPEModularMonitors2062971-001