Location
ExplanationWhattodo
RVPsyshigh/RVPsys
low
RVPmeanhigh/
RVPmeanlow
RVPdiahigh/RVPdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
SensorP1failedto
SensorP8failed
al.areaIP:Faultyordisconnected
sensor.
Checkthecableand
connections.
Replacethetransducer.
Sensor
param.
IP:Faultyordisconnected
sensor.
Checkthecableand
connections.
Replacethetransducer.
UACdisconnectal.areaInvasivepressurelineis
disconnected.
Checkthepatientstatus.
Checkconnections.
Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
UACsyshigh/
UACsyslow
UACmeanhigh/
UACmeanlow
UACdiahigh/UACdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
UVCsyshigh/
UVCsyslow
UVCmeanhigh/
UVCmeanlow
UVCdiahigh/UVCdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Zeroadj>100mmHg
param.
IPchannelzeroedtoover100
mmHgpressure.
Repeatthetransducer
zeroing.
Replacethesensor.
Replacethetransducer.
Replacethemodule.
Iftheproblempersists,
contactauthorizedservice
personnel.
Zeroed
param.
Zeroingwassuccessful.Noactionrequired.
Messageisautomatically
removedafter10seconds.