Location
ExplanationWhattodo
P1syshigh/P1sys
low
P1meanhigh/
P1meanlow
P1diahigh/P1dia
low
to
P8syshigh/P8sys
low
P8meanhigh/
P8meanlow
P8diahigh/P8dia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
P1zeroingfailedto
P8zeroingfailed
param.
Thechannelhasnotbeenzeroed
successfully.
Repeatthezeroing.
P7connectedal.areaThechannelhasbeen
connected.
Noactionrequired.
P8connectedal.areaThechannelhasbeen
connected.
Noactionrequired.
PAsyshigh/PAsys
low
PAmeanhigh/
PAmeanlow
PAdiahigh/PAdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
PR(ABP)high/PR(ABP)
low
PR(Art)high/PR(Art)
low
PR(Fem)high/
PR(Fem)low
PR(UAC)high/PR(UAC)
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Pressure
measurement
removed
al.areaTheacquisitionmodulehasbeen
removed.
Reconnectifnecessary.
PressureSensed
param.
Pressurehasbeensensedduring
zeroing.
Opentheventingstopcockto
air.
RAPsyshigh/
RAPsyslow
RAPmeanhigh/
RAPmeanlow
RAPdiahigh/RAPdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
2062971-001CARESCAPEModularMonitors479