ExplanationWhattodo
IdenticalPPmodulesal.areaTherearetwoormoreE-PP
modulesinthesystem.
RemoveallbutoneE-PP
module.
IdenticalPTmodulesal.areaTherearetwoormoreE-PT
modulesinthesystem.
RemoveallbutoneE-PT
module.
IPmoduleininvalid
slot[1,2,3,4]
al.areaTheinvasivepressuremoduleis
inthewrongmoduleslot.
Insertsingleparameter
pressuremoduleineitherslot
3or4oftheTram-Rac4A
housing.
IP'snotzeroedal.areaTheinvasivepressurelineshave
notbeenzeroed.
Performzeroing.
LAPsyshigh/LAPsys
low
LAPmeanhigh/
LAPmeanlow
LAPdiahigh/LAPdia
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
NoP1transducerto
NoP8transducer
al.areaNotransducerconnectedto
thechannelindicatedinthe
message,orthesensorisfaulty.
Connectatransducer.
Checkthecableand
connections.
Replacethesensor.
Replacethetransducer.
Iftheproblempersists,
contactauthorizedservice
personnel.
P1overrangetoP8
overrange
P1underrangetoP8
underrang
al.areaThemeasurementvalueisover
orunderrange,orthesensoris
faulty.
Checkthepatientstatus.
Checkthecables.
Rezerothetransducer.
Replacethesensor.
Replacethetransducer.
Replacethemodule.
Iftheproblempersists,
contactauthorizedservice
personnel.
P1standbytoP8
standby
param.
TheIPchannelhasbeensetto
standby.
Reactivatethechannelby
selectingActivateP1to
ActivateP8.
478CARESCAPEModularMonitors2062971-001