Messages
Message
Location
ExplanationWhattodo
●IdenticalPPmodules●al.areaTherearetwoormoreE-PP
modulesinthesystem.
●RemoveallbutoneE-PP
module.
●IdenticalPTmodules●al.areaTherearetwoormoreE-PT
modulesinthesystem.
●RemoveallbutoneE-PT
module.
●IPmoduleininvalid
slot[1,2,3,4]
●al.areaTheinvasivepressuremoduleis
inthewrongmoduleslot.
●Insertsingleparameter
pressuremoduleineitherslot
3or4oftheTram-Rac4A
housing.
●IP'snotzeroed●al.areaTheinvasivepressurelineshave
notbeenzeroed.
●Performzeroing.
●LAPsyshigh/LAPsys
low
●LAPmeanhigh/
LAPmeanlow
●LAPdiahigh/LAPdia
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●NoP1transducerto
●NoP8transducer
●al.areaNotransducerconnectedto
thechannelindicatedinthe
message,orthesensorisfaulty.
●Connectatransducer.
●Checkthecableand
connections.
●Replacethesensor.
●Replacethetransducer.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●P1overrangetoP8
overrange
●P1underrangetoP8
underrang
●al.areaThemeasurementvalueisover
orunderrange,orthesensoris
faulty.
●Checkthepatientstatus.
●Checkthecables.
●Rezerothetransducer.
●Replacethesensor.
●Replacethetransducer.
●Replacethemodule.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●P1standbytoP8
standby
●param.
TheIPchannelhasbeensetto
standby.
●Reactivatethechannelby
selectingActivateP1to
ActivateP8.
478CARESCAPEModularMonitors2062971-001