Cleaningandcare
Cleaningandcarepointstonote
Donotletliquidpoolaroundconnectionpins.Ifthisshouldhappen,blotdrywitha
soft,lint-freecloth.
Donotuseexcessivedryingtechniques,suchasoven,forcedheat,orsundrying.
Donotspraycleanerdirectlyonthedisplayscreen.
Neverconnectanydeviceorappliedparttoapatientuntilitisthoroughlydry.
Permitteddetergents
Water.
Mildsoap.
Permitteddisinfectants
Ethanol(max.99.7%byvolume).
Isopropylalcohol(max.60%byweight).
Chloramine(max.5%byvolume).
Glutaraldehyde(max.2%byvolume).
Phenol(max.2%byvolume).
Tartaricacid(75mgper100mlsolution).
Sodiumhypochlorite(max.5.25%byvolumemixedwithH
2
Oinratioof1:10).
Cleaningandcareinstructions
Settingthetouchscreenoffforcleaning
Youcansetthetouchscreenfeatureofffor30secondsatatimewhenyouneedto
cleanthescreen.Thecountdowntimerappearsonthescreen.
1.SelectMonitorSetup.
2.SelectTouchscreenOff.
3.Youcanenablethetouchscreenimmediatelybypressinganymonitorhardkeyor
theTrimKnob,orbyselectingCancelintheTouchscreenOffdisplay.
Cleaningnon-appliedparts,generalinstructions
Followtheseinstructionstocleanthemonitor,modules,displayscreensurfaces,
EEGheadbox,andothernon-appliedpartsunlessthereareseparatepart-specic
instructions.
1.Turnoffthepowertotheequipment.
2.Disconnecttheequipmentfromthepowersupply.
3.Removeallcablesandbatteries(ifapplicable)andclosebatterydoor(s).
4.Dampenasoftlint-freeclothwithoneofthepermitteddetergentsordisinfectants.