ECG
NOTE
TheECGSizesettingaffectsarrhythmiadetectionandheart
ratecalculationsensitivity.Normalwaveformsize/QRS
detectionsensitivityis1x.Size2xandgreaterincreases
theQRSdetectionsensitivity.Thismaybehelpfulforlow
amplitudeQRSwaveforms.Usewithcautionsincebaseline
artifactmaybedetectedasaQRScomplex.
Selectingthehemodynamicwaveformsweepspeed
NOTE
Thissettingadjuststhewaveformspeedforallofthe
hemodynamicparameters.
1.SelecttheHRparameterwindow.
2.SelectanumericvaluefromtheHemodynamicSweepSpeedlist.
Thesmallerthevalue,theslowerthesweepspeed.
PrintingallECGwaveforms
NOTE
PressingGraphinthetelemetrytransmitterwillstartthe
telemetrywaveformprinting.
1.SelecttheHRparameterwindow.
2.SelectAllECGWaveforms.
3.SelectPrintPage.
4.YoucanstopprintingbyselectingStopPrintingorCancelPrinting.
SelectingtheECGwaveformlter
Youcanselecthowthewaveformappearsonthedisplayandontheprintout.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelectalterfromtheWaveformFilterlist.Choicesare:
●Diagnostic:
■TRAM:0.05to100Hz.
♦TRAMwitha10-leadwirecable:Thewaveformlterisautomaticallyset
toDiagnosticandcannotbechanged.
■E-modulesandPDM:0.05Hzto150Hz.
●Monitoring:
■PDM,E-modules,TRAM:0.05Hzto32Hz(with50Hzpowerlinefrequency).
■PDM,E-modules,TRAM:0.05Hzto40Hz(with60Hzpowerlinefrequency).
■Telemetrytransmitters:0.05Hzto40Hz.Thewaveformlteris
automaticallysettoMonitoringandcannotbechanged.
●Moderate:
■PDM,E-modules,TRAM:0.05Hzto25Hz.
●Maximum:
■PDM,E-modules,TRAM:5Hzto25Hz.
160CARESCAPEModularMonitors2062971-001