Settingupthemonitorbeforeuse
CAUTION
Thewaveformautoscalingfeatureautomaticallyupdatesthe
displayfromthebestpossiblesignalamplitude.Alwaysmake
surethatthewaveformdisplayscaleiscorrectlyunderstood
anddoesnotleadtodelayedpatienttreatment.
SelectingthedisplaymodeforIPwaveforms
Youcanselecttheinvasivepressurewaveformstobeshownasindividualwaveforms,
orinacombinedview.
1.SelectMonitorSetup>ScreenSetup.
2.B850andB450withtheDoubleVideolicense:SelectScreen1orScreen2tab.
3.SelectUpperParameterArea.
4.SelectanoptionfromtheInvasivePressureWaveformslist:
Toviewindividualwaveforms,selectIndividual.
Tocombinethecurrentlydisplayedadjacentwaveforms(2to4),select
Combined.Thenewwaveformeldwillusethecombinedheightoftheoriginal
elds.
Tocombineuptofourwaveformsinoneeld,select4invP.Thenewwaveform
eldwillusetheheightoftwoupperparameterwindows.
Settingupasplitscreen
Youcansplitthewaveformareaintotwoparts.Thesplitscreenoptiondividesthe
screensothatyoucanviewgraphicand/ornumericdataofthechosenmeasurement
ontheleftwhilestillhavingwaveformsandparameterwindowsvisibleatthesame
time.
1.SelectMonitorSetup>ScreenSetup.
2.B850andB450withtheDoubleVideolicense:SelectScreen1orScreen2tab.
3.SelectSplitScreen.
4.Selectthetypeofsplitscreenyouneedfromthedropdownlist:
Off:nosplitscreen
STshowscurrentandreferenceQRScomplexesandSTtrends.
Spiro1isabasicviewofPatientSpirometrydata.
Spiro2isanenhancedviewofPatientSpirometrydata.
EEGshowstheEEGcompressedspectralarray(CSA).
AEPshowsthecurrentauditoryevokedpotentials(AEP)waveforms.
Minitrendshowsminitrendsbesidewaveforms.
Lockedalarmandparametersettings
Somecareunitandprolesettingscanbelocked.Clinicianscannotadjustthelocked
settingsfortheadmittedpatient.Thesesettingsareindicatedwithalocksymbol:
.
Thealarmandparametersettingsthatcanbelockedvarybyacquisitionmodule.
TheyaresetintheCareUnitSettingsandtheyarepasswordprotected.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.