ECG
NOTE
A6-or10-leadwirecablemaybeused.However,whenusing
a10-leadwirecable,donotprepareorconnectprecordial
leads2,3,4,or6.PositiontheCa/VaelectrodeintheC1/V1
positionandplacetheCb/VbelectrodeintheC5/V5position.
TheleadwirelabelfortheCa/VaandCb/Vbleadsarewhite
(IEC)orbrown(AAMI/AHA).
Ifyouareusingthe6–leadwirecablesfora12leadECGconnection,notethatthe
12RLcanbeusedforadultpatientsonly.
IEC
AAMI/AHA
Electrodeplacement
R(red)RA(white)Justbelowtherightclavicle.
L(yellow)LA(black)Justbelowtheleftclavicle.
Ca/C1(white)Va/V1(brown)
4
th
intercostalspace,rightsternalborder.
Cb/C5(white)Vb/V5(brown)LeftanterioraxillarylineatC4/V4level.
N(black)
RL(green)
Lowerrightedgeoftheribcage.
F(green)
LL(red)Lowerleftedgeoftheribcage.
10–leadECGelectrodeplacementforcardiacmonitoring
152CARESCAPEModularMonitors2062971-001