EasyManua.ls Logo

GE CARESCAPE - NMT measurement basics; NMT measurement description

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Neuromusculartransmission
1.Regionalblockadapter
2.Sensorcable
3.Sterileneedle
Preparingforlocalnerveorplexusstimulation
1.ConnecttheE-NMTregionalblockadaptertothesensorcable.
2.Connectadisposableneedleandsyringesettotheadapter.
3.PresstheStart-upmodulekeyandselectasuitablecurrentfromtheCurrentmA
listthatappearsonscreen.YoucanalsodothisbyselectingtheNMTparameter
window>RegionalBlocktab>CurrentmA.
4.StartstimulationbyselectingStart.
5.Whentheneedleapproachesthemotornerve,reducethecurrent.Thepulse
widthis40msandsomewhathighercurrentsmaybeneededthaninother
similarsystems.
6.Whenready,stopthestimulationbypressingtheStopContinuemodulekey,or
byselectingtheNMTparameterwindow>RegionalBlock>Stop.
NMTmeasurementbasics
NMTmeasurementdescription
TheGENMTmeasurementdevicesareusedformonitoringtherelaxationofthe
patientandregionalblockstimulationfornervelocation.
Neuromusculartransmissionisthetransferofamotornerveimpulseoverthe
neuromuscularjunction.TheGENMTdevicesdeliverstimulatingelectricalpulsestoa
motornerveandthemuscleresponsetothesestimulationsismeasured.
Twoelectrodesareneededforelectricalstimulationofaperipheralnerve.The
resultingresponsecanbemeasuredwitheithertwoelectrodesandaMechanoSensor,
whichmeasuresmovementsbetweenthumbandindexnger,oranElectroSensor
usingthreerecordingelectrodes.
Themonitorsearchesforthestimuluscurrentneededtoactivateallthebersofthe
stimulated(recorded)muscles.Thesearchbeginswitha10mAstimulusandthe
responseismeasured.Thecurrentisincreasedinstepsof5mAuntiltheincrease
inthecurrentnolongerincreasestheresponse.Thismaximalcurrentisthen
automaticallyincreasedby15%,resultinginthesupramaximalcurrent.
Ifthesupramaximalcurrentisnotfoundortheresponseistooweakforsearchinga
supramaximalcurrent,thecurrentissetto70mA.
MechanoSensorandkinemyography(KMG)
MeasurementwithMechanoSensorisbasedonkinemyographyprinciple.Two
electrodesareplacedontheulnarnerve,andtheinnervatingnerveisstimulated.
ThecoreoftheMechanoSensorisastripofpiezoelectricpolymer,housedinsidethe
sensor.Whenthispiezoelectricmaterialchangesshapeduetomusclecontraction
asaresultofnervestimulation,theelectricalchargeinthematerialisredistributed,
causinganelectronowtobalancethecharges.Thisowcanbemeasuredasa
potentialchange,whichisproportionaltotheamountofmaterialbending.
368CARESCAPEModularMonitors2062971-001