Viewingothermonitoredpatients
Automaticviewofremotebedsinalarm
Automaticviewingofremotealarmsisalicensedfeature.
Youcansetthemonitortoautomaticallynotifyyouwithanalarmmessageorwitha
bed-to-bedwindowwhenselectedremotepatientbedsgointoanalarmcondition.
Allautomaticallyviewablealarmingbedsdisplayinorderfromthehighesttolowest
alarmpriorityandfromthenewesttooldestalarms.
Youcancongurehowthemonitornotiesyouofaremotepatientbedalarm
condition,andwhichremotepatientalarmprioritylevelsyouwantnoticationof.You
candothisforindividualbeds,orforallremotebedsofaselectedcareunitatonce.
Selectingthealarmnoticationtype
1.SelectData&Pages.
2.SelectOtherPatients.
3.SelecttheReceiveAlarmstab.
4.SelectacareunitfromtheUnitlist.
5.Selectapatientbedfromthedisplayedlist.
6.SelectthetypeofalarmnoticationdesiredfromtheAlarmNoticationlist:
●Off:Remotealarmnoticationisturnedoff.
NOTE
Allnewmonitorsappearonthelistautomaticallywith
thenoticationsettingOff(default).
●Message:Remotealarmmessagesdisplayinthealarmarea.Atanytime,
youcanselectaremotealarmmessagetoopenabed-to-bedwindowand
viewtheremotepatient’sdata.
●AutoView:Bed-to-bedwindowopensimmediatelyifnootherprocedureor
setupwindowiscurrentlyopen.Otherwise,aremotealarmmessagedisplays
inthealarmmessagearea.Toopenthebed-to-bedwindowandviewthe
remotepatient’sdata,closethecurrentlyopenmenu,orselecttheremote
alarmmessage.
●AutoViewAlways:Immediatelyclosesanyopenprocedureorsetupwindows
andopensabed-to-bedwindowwiththeremotepatient’sdata.
Selectingthenotifyingalarmprioritylevel
1.SelectData&Pages.
2.SelectOtherPatients.
3.SelecttheReceiveAlarmstab.
4.SelectacareunitfromtheUnitlist.
5.Selectapatientbedfromthedisplayedlist.
444CARESCAPEModularMonitors2062971-001