Interfacingwithperipheraldevices
Devicetype
Waveforms
1
Parameter
windows
TrendsAlarm
broadcast
Printouts
Real-timedata
tocentral
station
2
Pulse
oximeters
NoYesYesYesYesNo
Transcuta-
neousmoni-
tors
NoYesYesYesYesNo
Ventilators
YesYesYesYesYes
Limited
Gasanalyzers
NoYesYesYesYesYes
Continuous
cardiacoutput
NoYesYesYesYesNo
Anesthesia
machines
YesYesYesYesYes
Limited
POCbloodgas
monitors
NoNo
3
YesNoNoNo
1
UnityNetworkInterfaceDevice(ID)supportsonlydigitalwaveforms.
2
CICProversion4.0.7andsubsequent.CARESCAPECentralStationsoftwareversion1.0andsubsequent.
3
POCbloodgasmonitordataisdisplayedinData&Pages>LaboratoryData.
Peripheraldevicedatapresentationandmenus
Thedatadisplayedintheparameterwindowmayvarywitheachdevice.Notall
menuoptionsareavailablewithalldevices,andsomemenuoptionsarenotavailable
withinterfaceddevicesatall.Ventilators,gasanalyzers,continuouscardiacoutput
devices,andanesthesiamachinesarecapableofsendinganumberofparameters
tothemonitor.Insomecases,notallaresupported,orthemonitor’ssoftware
packagemaydeterminethesetofparameters.SeetheUnityNetworkInterface
Device(ID)Operator'sManualandtheperipheraldevicemanufacturer’smanualsfor
moreinformation.
Pulseoximeters
TheSpO
2
parameterwindowdisplaysasaturationvalueandapulserate.
TheExtindicatesthedataisfromanexternalsource.Thereisnoassociated
waveformorparametermenu.OnlyoneSpO
2
measurementfromanexternal
sourcecanbeactivelydisplayedonthemonitor.
SvO
2
TheSvO
2
parameterwindowdisplaysSvO
2
indexdataandasignalquality
indicator(None,*,**,***).Forreliablesaturationvalues,thesignalstrength
indicatorshouldbehigherthanoneasterisk.ThemeasurementrangeofSvO
2
is
from0to100%.TheExtindicatesthedataisfromanexternalsource.
TranscutaneouspO
2
/pCO
2
TheTCparameterwindowdisplaysapCO
2
value,pO
2
value,sitetemperature
value,andprobepowervalue.TheExtindicatesthedataisfromanexternal
source.Thereisnoassociatedwaveform.OnlyoneTCmeasurementfroman
externalsourcecanbeconnectedandactivesimultaneously.
Spirometrydatafromventilatorsandanesthesiamachines
2062971-001CARESCAPEModularMonitors451