Messages
Message
Location
ExplanationWhattodo
Measurementoff
param.,
wavef.
ECGleadsnotconnectedtothe
patient.
ConnecttheECGleadsto
thepatienttostartthe
impedancerespiration
measurement.
Relearning
param.,
wavef.
Thepatient’sbreathingpattern
isbeingrelearnedoraleadhas
beenchanged.
Waituntilthemessage
disappears.
RemoveoneECG
module
al.areaTherearetwoECGmodulesin
thesystem.
RemoveoneECGmodule.
RR(Imped)High/RR
(Imped)Low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Smallrespcurve
param.
Signalamplitude<0.4ohm.Checkthepatientstatus.
Checkelectrodeplacement.
Changeormoveelectrodes.
MessagesrelatedtoSpO
2
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
report=reportview
wavef.=waveformarea
Message
Location
ExplanationWhattodo
Artifact
param.
Artifactdetected.Checksensorcontact.
Reposition/replacesensor.
Requestthepatienttoremain
still.
Checkdevice
param.
Theacquisitionmodulehas
failed.
Replacetheacquisition
module.
CheckExtdevice
param.
Interfaceddeviceshouldbe
checked.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
Checkprobe
param.
ThereisnodetectableSpO
2
signal,thesensorisfaultyoris
detachedfromthepatient.
Checkthesensorand
connections.
2062971-001CARESCAPEModularMonitors469