Messages
Message
Location
ExplanationWhattodo
CheckSpO2probe/
CheckSpO2(2)probe
al.areaThereisnodetectableSpO
2
signal,thesensorisfaultyoris
detachedfromthepatient.
Checkthesensorand
connections.
Connecting
param.
Connectiontotheinterfaced
deviceisbeingestablished.
Noactionrequired.
Faultyprobe
param.
Thesensorhasfailed.Replacethesensor.
IdenticalSpO2
modules
al.areaTherearetwoormoreidentical
SpO
2
modulesinthesystem.
RemoveallbutoneSpO
2
module.
Incompatibleprobe
param.
Thesensorisnotcompatible.Replacethesensor.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
Iftheproblempersists,
contactauthorizedservice
personnel.
IncompatibleSpO2
probe/Incompatible
SpO2(2)probe
al.areaThesensorisnotcompatible.Replacethesensor.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
Iftheproblempersists,
contactauthorizedservice
personnel.
Interference
param.
Themeasurementisdisturbed.Checkthesensor.
Learning
param.
SpO
2
algorithmisinlearning
phase.
Noactionrequired.
Lowbattery
param.
Theinterfaceddevicehaslow
batterycharge.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
Lowperfusion
param.
Lowperfusionatthe
measurementpoint.
Checkthesensorandsensor
positioning.
Relocatethesensortoa
bettermeasurementsite,if
possible.
Makesurethepatientisnot
shivering.
Lowsignal
param.
Thequalityofthesignalis
questionable.
Checkthesensorplacement
andthepatientstatus.
Lowsignalquality
param.
Thequalityofthesignalis
questionable.
Checkthesensorandsensor
positioning.
Relocatethesensortoa
bettermeasurementsite,if
possible.
Makesurethepatientisnot
shivering.