Messages
Message
Location
ExplanationWhattodo
●CheckSpO2probe/
CheckSpO2(2)probe
●al.areaThereisnodetectableSpO
2
signal,thesensorisfaultyoris
detachedfromthepatient.
●Checkthesensorand
connections.
●Connecting
●param.
Connectiontotheinterfaced
deviceisbeingestablished.
●Noactionrequired.
●Faultyprobe
●param.
Thesensorhasfailed.●Replacethesensor.
●IdenticalSpO2
modules
●al.areaTherearetwoormoreidentical
SpO
2
modulesinthesystem.
●RemoveallbutoneSpO
2
module.
●Incompatibleprobe
●param.
Thesensorisnotcompatible.●Replacethesensor.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●IncompatibleSpO2
probe/Incompatible
SpO2(2)probe
●al.areaThesensorisnotcompatible.●Replacethesensor.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Interference
●param.
Themeasurementisdisturbed.●Checkthesensor.
●Learning
●param.
SpO
2
algorithmisinlearning
phase.
●Noactionrequired.
●Lowbattery
●param.
Theinterfaceddevicehaslow
batterycharge.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Lowperfusion
●param.
Lowperfusionatthe
measurementpoint.
●Checkthesensorandsensor
positioning.
●Relocatethesensortoa
bettermeasurementsite,if
possible.
●Makesurethepatientisnot
shivering.
●Lowsignal
●param.
Thequalityofthesignalis
questionable.
●Checkthesensorplacement
andthepatientstatus.
●Lowsignalquality
●param.
Thequalityofthesignalis
questionable.
●Checkthesensorandsensor
positioning.
●Relocatethesensortoa
bettermeasurementsite,if
possible.
●Makesurethepatientisnot
shivering.
470CARESCAPEModularMonitors2062971-001