Messages
Message
Location
ExplanationWhattodo
●Motiondetected
●param.
Patientmovementdetected.
●Repositionsensor.
●Noextdevice
●param.
Interfaceddeviceshouldbe
checked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Noprobe
●param.
Sensorisnotconnectedtothe
acquisitionmodule.
●Checkconnectionbetween
thesensorandtheacquisition
module.
●Nopulse
●param.
Pulsesignalispoor.●Tryothermeasurementsites.
●NoSpO2probe/No
SpO2(2)probe
●al.areaSensorisnotconnectedtothe
acquisitionmodule.
●Checkconnectionbetween
thesensorandtheacquisition
module.
●NoSpO2pulse/No
SpO2(2)pulse
●al.areaPulsesignalispoor.●Tryothermeasurementsites.
●PR(SpO2)high/
PR(SpO2)low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Probeoff
●param.
Thesensormaybedefective.●Checkthepatientstatus.
●RepositiontheSpO
2
sensor.
●ReplacetheSpO
2
sensor.
●Pulsesearch
●param.
Defectiveordamagedsensor
orcable.Sensorisoffof
thepatient.Detectionofa
repeatablepulsehasstopped.
●Checkthesensorandcable.
●Repositionorreplacesensor.
●SpO2alarmsetup
changed
●al.areaAcquisitionmodulehasbeen
movedtoanotherchanneland
thealarmlimitsorthealarm
on/offstatusofthisnewchannel
aredifferentfromtheprevious
channel.
●Youcanremovethemessage
fromthescreenwiththe
pauseaudiokey.
●SpO2faultyprobe/
SpO2(2)faultyprobe
●al.areaThesensorhasfailed.●Replacethesensor.
●SpO2high/SpO2low
●SpO2(2)high/SpO2(2)
low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●SpO2measurement
removed
●al.areaAsecondarySpO
2
source
becomestheprimarysourceas
aresultofremovingPDM,PSM,
orTRAM.
AlsowhentheTRAMcableis
disconnected.
●Connectamoduletothe
samechannel.
●ConnecttheTRAMmodule
andcable.
●SpO2probeoff
/SpO2(2)probeoff
●al.areaThengerorearlobemaybe
toothinorthesensorisoffthe
patient.
●Checkthepatientstatus.
●RepositiontheSpO
2
sensor.
●ReplacetheSpO
2
sensor.
2062971-001CARESCAPEModularMonitors471