Message
Location
ExplanationWhattodo
Motiondetected
param.
Patientmovementdetected.
Repositionsensor.
Noextdevice
param.
Interfaceddeviceshouldbe
checked.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
Noprobe
param.
Sensorisnotconnectedtothe
acquisitionmodule.
Checkconnectionbetween
thesensorandtheacquisition
module.
Nopulse
param.
Pulsesignalispoor.Tryothermeasurementsites.
NoSpO2probe/No
SpO2(2)probe
al.areaSensorisnotconnectedtothe
acquisitionmodule.
Checkconnectionbetween
thesensorandtheacquisition
module.
NoSpO2pulse/No
SpO2(2)pulse
al.areaPulsesignalispoor.Tryothermeasurementsites.
PR(SpO2)high/
PR(SpO2)low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Probeoff
param.
Thesensormaybedefective.Checkthepatientstatus.
RepositiontheSpO
2
sensor.
ReplacetheSpO
2
sensor.
Pulsesearch
param.
Defectiveordamagedsensor
orcable.Sensorisoffof
thepatient.Detectionofa
repeatablepulsehasstopped.
Checkthesensorandcable.
Repositionorreplacesensor.
SpO2alarmsetup
changed
al.areaAcquisitionmodulehasbeen
movedtoanotherchanneland
thealarmlimitsorthealarm
on/offstatusofthisnewchannel
aredifferentfromtheprevious
channel.
Youcanremovethemessage
fromthescreenwiththe
pauseaudiokey.
SpO2faultyprobe/
SpO2(2)faultyprobe
al.areaThesensorhasfailed.Replacethesensor.
SpO2high/SpO2low
SpO2(2)high/SpO2(2)
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
SpO2measurement
removed
al.areaAsecondarySpO
2
source
becomestheprimarysourceas
aresultofremovingPDM,PSM,
orTRAM.
AlsowhentheTRAMcableis
disconnected.
Connectamoduletothe
samechannel.
ConnecttheTRAMmodule
andcable.
SpO2probeoff
/SpO2(2)probeoff
al.areaThengerorearlobemaybe
toothinorthesensorisoffthe
patient.
Checkthepatientstatus.
RepositiontheSpO
2
sensor.
ReplacetheSpO
2
sensor.
2062971-001CARESCAPEModularMonitors471