Systemintroduction
4.Ethernet:ForconnectingtheMCandIXnetworks.TheMCNetworkestablishes
communicationandallowspatientdatatobesenttoanoptionalCICProClinical
InformationCenterorCARESCAPECentralStation.TheIXNetworkprovidesaccess
forexampletotheMUSEserver,Citrixserver,andIXprinters.
5.Tram-NetandePort:ForconnectingthePDM,E-moduleFrame,andTram-Rac
housing.
6.RS232:Forconnectingthetouchscreendisplay.
7.DVI-1supportsadigitaldisplayandclonedanalogdisplay.DVI-2supportsonly
onedigitaldisplay.DVI-3(optionalthirdvideo)supportsadigitaldisplayanda
clonedanalogdisplay(iPanelapplicationonly).
8.Groundplug:Providesequipotentiality.
B850moduleframesF7andF5
WARNING
Donottouchtheelectricalconnectorlocatedwithinthe
modulehousingorframe.
WARNING
—B850:ELECTRICSHOCK—DonotusetheF7Framefor
standaloneuse.VentilationholesontheF7E-moduleFrame
willbecoveredonlyifinstalledwithinanAisys,Avance,or
Aespireanesthesiamachine.
1.Debrillatorsynchronization
F5andF7E-moduleFramesprovideaninterfacebetweenthemonitorandE-modules.
Framesallowadditionalparameterstobemonitored.
TheF5FramehasvemoduleslotsthatsupportallE-moduleacquisitionmodules.It
supportsbothPDMandPSMwithaslidemount.
TheF7Framehassevenmoduleslots,butitdoesnothaveaslidemountforthePDM
orPSM.ThePSMmodulemaybeinterfacedtotheF5orF7Framewithacablewhen
theModuleBusAdapterorInterfaceModuleforPSMisused.