Messages
Message
Location
ExplanationWhattodo
●SvO2high/SvO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●SvO2measurement
removed
●al.areaAcquisitionmodulehasbeen
removed.
●Connectthemoduleif
youwanttorestartthe
measurement.
●SvO2notcalibrated●al.area,
param.
Theopticalmoduleisconnected
tothemonitorandthecatheter
hasnotbeencalibrated.
●Performinvivocalibration.
●SvO2outofrange
●param.
Valuesareaboveorbelowthe
rangethemonitorcanprocess.
●Checkcablesand
connections.
●Repeatinvivocalibration.
●Changetheopticalmodule.
●Ifthecatheterisfaulty,
replaceit.
●Ifthecableormoduleis
broken,contactauthorized
servicepersonnel.
●SvO2signalpoor●al.area,
param.
Thereissignalpulsation,the
catheteristouchingthewall,
orthereisanintensityshiftin
signalqualitylevel.
●Flushthecatheter.
●Checktheopticalmoduleand
connections.
●Checkthecatheter
placement.
●SvO2temperror
●al.areaThetemperatureoftheoptical
moduleisoutofrangeformore
than10minutes.
●Checktheopticalmoduleand
connections.
●Replacetheopticalmodule.
●Contactauthorizedservice
personnel.
●Wait,initializing
●param.
Opticalmodulesignalisbeing
processed.
●Waituntilthemessage
disappears.
●Warmingup
●param.
Theopticalmoduleiswarming
up.
●Waituntilthemodulehas
warmedupandthemessage
disappears.
Messagesrelatedtogasesmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seetheCARESCAPEModular
MonitorsSupplementalInformationManual.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●report=reportview
●wavef.=waveformarea
488CARESCAPEModularMonitors2062971-001