Messages
Message
Location
ExplanationWhattodo
●Agentmixture
●al.area,
param.
Mixtureofhalogenatedagents
isdetected.
●Checktheventilatorand
agentvaporizersettings.
●Apnea(CO2)
●al.area,
wavef.
Nobreathingdetected.●Checkthepatientstatus.
●Checktheventilatorand
breathingstatus.
●Apneadeactivated
●param.
Anewcasehasjustbeen
started/anewpatientadmitted,
orthemeasurementhasjust
beenstartedandtheapnea
alarmisnotactiveyet.
●Wait.Themessage
disappearsafterthemonitor
detects3breathsduringthe
lastminute(E-modules),or
therespirationrateis≥3per
minute(CAPNOSTAT,Dual
CO2,CapnoFlex).
●Calibratesensor
●param.
Anewsensorisintroduced
orthereisasignaldriftinthe
electronics.
●Performsensorcalibration.
●Calibratesystem
●param.
Interfaceddeviceshouldbe
calibrated.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Calibrating
●param.
Calibrationisinprogress.●Noactionrequired.
●Calibratinggas
sensor
●wavef.Themeasurementhasbeen
startedandcalibrationisin
progress.
●Waituntilthecalibrationis
completedandthemessage
disappears.
●Calibrationerror
●param.
Unsuccessfulcalibration.●Authorizedservicepersonnel
shouldrepeatthecalibration
procedure.
●Check/calibr.adapter●al.area,
param.
Theadapterismissing,
obstructed,orofadifferent
typethaninthelastcalibration.
●Checkthatthereisanadapter
inthesystem.
●Checktheadapterfor
blockages.
●Calibratetheadapter.
●Checkextdevice
●param.
Interfaceddeviceshouldbe
checked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Checksamplegasout●al.areaThewatertrapisnotconnected,
thesamplegasoutowis
blocked,orthereisaleakinside
themodule.
●Checkwatertrapconnection.
●Removetheblockagefrom
thesamplegasoutlet.
●Changemoduleifneeded.
●Checksampleline●al.area,
param.
Thesamplelineisdisconnected,
orofadifferenttypethanthe
lastoneused.
●Reconnectthedisconnected
sampleline.
●Calibratethesampleline.
2062971-001CARESCAPEModularMonitors489