GE CARESCAPE - Page 490

GE CARESCAPE
548 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Messages
Message
Location
ExplanationWhattodo
●Checkwatertrap
andsamplegasout.
PressNormalScreen
tocontinue.
●wavef.Thewatertrapisnotconnected,
thesamplegasoutowis
blocked,orthereisaleakinside
themodule.
●Checkwatertrapconnection.
●Removetheblockagefrom
thesamplegasoutlet.
●Changemoduleifneeded.
●CheckWaterTrap●al.area,
param.
Watertrapconnectionisnot
correct,orthereisaleakinthe
samplinglineinsidethemodule.
●Checkthepatientstatus.
●Checkthewatertrapandits
connection.
●Continuousblockage,
checksampleline.
Restartpump.
●wavef.Nasalcannula,moisturelter,
sidestreamadaptertube,or
exhaustlineisblocked.
●Checkallpartsofthe
equipmenttopatient
connectionandremove
anyblockages.
●Ifnecessary,changetheline.
●Continuousblockage.
Checksampleline
andwatertrap.
●wavef.Thegassamplinglineisblocked
orthewatertrapisoccluded.
●Changesamplinglineand
watertrap.
●EtAAhigh
●EtAAlow
whereAA=Hal,Enf,Iso,
SevorDes
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●EtCO2high
●EtCO2low
●EtN2Ohigh
●EtN2Olow
●EtO2high
●EtO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Ext:Apnea
●al.areaNobreathingdetected.●Checkthepatientstatus.
●Checktheventilatorandthe
breathingstatus.
●Checktheinterfaceddevice.
●Extnotsupported
●param.
Interfaceddeviceisnot
supported.
●SeetheUnityNetwork
InterfaceDevice(ID)
Operator'sManualfor
supporteddevices.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●FailureinAgentID
●param.
Anunknownagentorthreeor
moreagentsdetected.Vaporizer
maycontainamixtureofagents.
●Flushthebreathingcircuit
withO
2
ush(O
2
+,100%O
2
).
●Emptythevaporizerandrell
fromanunopenedcontainer.
490CARESCAPEModularMonitors2062971-001