Messages
Message
Location
ExplanationWhattodo
●FiAAhigh
●FiAAlow
whereAA=Hal,Enf,Iso,
SevorDes
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●FiCO2high
●FiCO2low
●FiN2Ohigh
●FiN2Olow
●FiO2high
●FiO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Gasmeasurements
removed
●al.areaAcquisitionmodulehasbeen
removed.
●Connectthemoduleif
youwanttorestartthe
measurement.
●Gasmodulestandby●al.areaAcquisitionmodulehasshut
downthepump.
●Noactionrequired.
●Gasmodulestandby.
Touchanybutton/key
toactivate.
●wavef.Acquisitionmodulehasshut
downthepump.
●Toexitthestandbymode,
touchanybuttonorkey.
●Noactionrequired.
●Identicalgasmodules●al.areaTherearetwoormoreE-modules
withgasmeasurementinthe
system.
●RemoveallbutoneE-module
withgasmeasurement.
●Incompatiblegas
module
●al.areaThemoduleisnotcompatible.●Replacewithcompatible
gasmodule.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Incompatiblesensor
●param.
Thesensorisofwrongtype.●Checkandchangethesensor
toacorrecttype.Seethe
CARESCAPEModularMonitors
SupplementalInformation
Manual.
●InterfaceFailed
●param.
InterfaceddeviceandUnity
NetworkInterfaceDevice(ID)
shouldbechecked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Lowgassampleow●al.areaSampleowislessthan80%of
themodule’snominalowvalue.
Thiscanhappenifnebulized
medicationsaregivenwithout
disconnectingthesampleline.
●Checkthesampleline.
●Iftheproblempersists,
contactauthorizedservice
personnel.
2062971-001CARESCAPEModularMonitors491