Airwaygases
●Makesurethatthegassamplinglineisproperlyconnectedtothewatertrapand
thewatertrapisproperlyconnectedtotheairwaygasmodule.Gasleaksinthese
connectionsmaydilutethegassamplefromthepatientcircuit,thusresultingin
erroneousgasreadings.Duringnormaloperation,allsampledgasowsoutofthe
samplegasoutlet.Roomairisusedasreferencegasfortheoxygenmeasurement
anditismixedwiththesampledgas.Thesampledgasisdilutedbyroomairso
thatthefractionofroomairintheexhaustgasisabout20%.
●Compactairwaymodules:ThemessageSamplelineblockedmayresultifyou
attachthesamplinglinetothewatertrapafterthemonitorhascompletedthe
self-checkforthemodule.Attachthesamplinglinestothewatertrapbefore
turningonthemonitor.
Airwaygasesmeasurementsetup
Airwaygasesequipmenttopatientconnectionswith
CARESCAPErespiratorymodules
1.CARESCAPErespiratorymodule
2.Gassample,gassamplinglineconnectoronthewatertrap
3.Gassamplingline
4.Gassamplinglineconnectorontheairwayadapter;placetheconnectorupwards
5.Airwayadapterwithsamplinglineconnector
6.Heatandmoistureexchangerwithlter(HMEF)(optional)
284CARESCAPEModularMonitors2062971-001