15
Mixedvenousoxygensaturation
(SvO₂)
SvO
2
compatibilitylimitations
AllfeaturesofthismeasurementareavailablewithallBx50monitors.
Fordetailedinformationregardingmodule,monitor,andaccessorycompatibility,see
theCARESCAPEModularMonitorsSupplementalInformationManual.
SvO₂safetyprecautions
SvO
2
warnings
WARNING
Allinvasiveproceduresinvolveriskstothepatient.Useaseptic
technique.Incorrectuseofthecathetercanleadtovessel
perforation.Followcathetermanufacturer'sinstructions.
WARNING
Whenusinganelectrosurgeryunit,notethatthe
measurementcablesdonotincorporatemeanstoprotect
againstburnsincaseofadefectiveESUreturnelectrode.To
avoidburnsatthemonitormeasurementsites,ensurethe
following:
PropercontactoftheESUreturnelectrodetothepatient.
ESUreturnelectrodeneartheoperatingarea.
Measurementelectrodes,leadwiresandprobesfarfrom
thesurgicalsiteandtheESUreturnelectrode.
WARNING
Theuseofsuchdyesthatusuallychangethepatient
bloodpigmentationmayleadtofaultyoxygensaturation
measurementvalues.
SvO
2
measurementlimitations
TheSvO
2
measurementisnotintendedforneonatalpatients.
2062971-001CARESCAPEModularMonitors275