Pulseoximetry
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2tab.
3.AdjustthevolumewiththeBeatVolumearrows.
Variablebeattone
YoucancongureavariablebeattonethroughMonitorSetup>DefaultSetup>Care
UnitSettings>Parameters>VariableBeatTone.Thissettingispasswordprotected.
IfitsettoAllbeatsources,theSpO
2
saturationaffectsallbeepsoundsincludingECG
andIPwhentheSpO
2
measurementisavailable:beepfrequencychangesaccording
toincreasinganddecreasingSpO
2
values.IfthesettingissettoOnlySpO2,otherbeep
soundsarenotaffectedbythechangingSpO
2
values.
Formoreinformation,seetheCARESCAPEModularMonitorsSupplemental
InformationManual.
MasimoSETdataaveragingandupdating
ForMasimoSETtechnology,whenusingthedefaultaveragingtimeof8seconds,
thereisamaximumdata-averagingsignalprocessingtimeof10secondsfromreal
timeplusanadditionaldelayof2secondstoupdatethedisplayedwaveform.Audible
alarmsaredelayeduntilanalarmlimitviolationoccursforatleast5seconds.
SelectingtheSpO
2
averagingtime
NOTE
PSM,E-MASIMO,andPDMandTRAMwithMasimotechnology
andMasimosensorsonly.TheprimarySpO
2
measurement
only.
YoucanhaveanaverageoftheSpO
2
measurementonscreeninsteadofthebeatto
beatvalues,andyoucanselecthowmanysecondsareusedforthisaveraging:2s,
4s,8s,10s,12s,14s,or16s.
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2tab.
3.SelectthenumberofsecondsfromtheAveraginglist.
SelectingtheMasimoSpO
2
sensorsensitivitylevel
NOTE
SpO
2
moduleswithMasimotechnologyandMasimosensors
only.
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2tab.
3.SelecttheappropriateSensitivityradiobutton:
●UsetheNormalsensitivitysettingfornormalpatientmonitoringpurposes.
●UsetheMaximumsensitivitysettingforimprovedpoorperfusionperformance
andforfastertrackingofrapidSpO
2
saturationchanges.
UsingtheMaximumsensitivitysettingdelaystheSpO2probeoffdetectionalarm.
210CARESCAPEModularMonitors2062971-001