ECG
AboutthereferenceQRS
TRAMmodulessavetheinitialreferenceQRSafter64beatshavebeendetected.With
TRAMonereferenceQRScomplexcanbestoredmanually.
PDMandE-modulessavetheinitialreferenceQRSuptothreeminutesaftertheECG
measurementisstarted.WithPDMorE-modulesuptosixadditionalreferenceQRS
complexescanbestoredmanually.
EachQRSreferenceisidentiedwiththedateandtimestamp.
IfanewreferenceQRScomplexissavedandthereisnoroomforanotherreference
complex,thentheoldestmanualorautomaticreferencecomplexiserased.
SavingareferenceQRSmanually
Thisselectionisnotavailablewithcombinationmonitoring.
YoucannotsaveareferenceQRSmanuallyuntilaninitialreferenceQRScomplexhas
beensaved,orifSTAnalysisisdisabled.
1.SelecttheSTparameterwindow.
2.SelectRealtimeView.
3.SelectSaveReference.
ThecurrentQRSbecomesthenewreferenceQRS.
AutomaticsavingofreferenceQRScomplexes
TheQRSreferenceisnotsavedduringanSTalarmcondition,onlytheSTsnapshot
issaved.
IfanewreferenceQRScomplexissavedandthereisnoroomforanothercomplex,
thentheoldestmanualorautomaticcomplexiserased.Youmaywanttomanually
eraseQRScomplexestoavoidautomatederasing.
AreferenceQRScomplexissavedautomaticallywheneveryoudooneofthefollowing:
●ChangetheVaorVblead.
●ChangetheSTpointmanually.
SelectingasavedreferenceQRScomplexfordisplay
ThisselectionisnotavailablewithcombinationmonitoringorTRAM.
YoucanselectanddisplayasavedreferenceQRSforSTanalysis.
1.SelecttheSTparameterwindow.
2.SelectRealtimeView.
3.SelectasavedreferencefromtheReferenceQRSlist.
ErasingareferenceQRS
ThisselectionisnotavailablewithcombinationmonitoringorTRAM.
YoucannoterasetheinitialreferenceQRS.
1.SelecttheSTparameterwindow.
2.SelectRealtimeView.
2062971-001CARESCAPEModularMonitors185