Trends
4.SelectavaluefromtheMinitrendLengthlist.
The1minuteand2minuteminitrends(high-resolutionlicense)areupdated
every2seconds.
The5minuteand10minuteminitrendsareupdatedevery10seconds.
Otherlengthsareupdatedonceaminute.
Selectinghigh-resolutioncontentstominitrend
Ifthehigh-resolutionlicenseisenabledandthe10minwideis1minutenormalor
2minuteswide,youcanselectCO
2
andimpedancerespiration(10Hz)Compressed
WaveformminitrendsorRespRateminitrends.
1.SelectMonitorSetup>ScreenSetup.
2.B850andB450withDoubleVideolicense:SelectScreen1orScreen2tab.
3.SelecttheSplitScreentab.
4.SelectMinitrendfromtheShowlist.
a.SelectRespRatefromtheCO2MinitrendorfromtheImp.RespMinitrendlist.
b.SelectCompressedWaveformfromtheCO2MinitrendorfromtheImp.Resp
Minitrendlist.
Removingminitrendfromthescreen
1.SelectMonitorSetup>ScreenSetup.
2.B850andB450withtheDoubleVideolicense:SelectScreen1orScreen2tab.
3.SelecttheSplitScreentab.
4.SelecttheNonefromtheShowlist.
Timechangeduringapatientcase
CAUTION
Systemtimechangeswillresultintimedifferencesbetween
storedandrealtimedata.
Timeadjustmentisallowedduringapatientcaseifthemonitorisconguredtothe
CARESCAPENetwork.Whenthetimeisadjusted,themonitorshiftsthetimestampsof
continuoustrenddataandofdiscretedata,exceptNIBPandwedgemeasurements.
Aftertimeadjustmentcontinuousanddiscretedatacannotbecomparedtoeach
otherbecausetheirtimestampsnolongermatch.
2062971-001CARESCAPEModularMonitors423