EasyManua.ls Logo

GE CARESCAPE - D-lite(+);Pedi-lite(+) flow sensor; Patient Spirometry parameters

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PatientSpirometry
D-lite(+)/Pedi-lite(+)owsensor
1.Fortotalpressuremeasurement
2.Forstaticpressuremeasurement
3.Forgassamples
Thevelocityofgasowisobtainedwhenthedynamicpressureismeasuredbythe
twohollowtubes(1and2).Oninspiration,gasmovesfromtheanesthesiamachineor
ventilatortothepatientbypoint1whichmeasuresthetotalpressure,andatthesame
timethepressureattube2ismeasuredasthestaticpressure.Thestaticpressureat
2issubtractedfromthetotalpressureat1togivethedynamicpressure.Dynamic
pressureisproportionaltothevelocityofgasow.D-lite(+)/Pedi-lite(+)isdesignedto
workinbothdirections;duringexpirationtheprocessisreversed.Gassamplesfor
CO
2
,O
2
,N
2
Oandanestheticagentmeasurementsaretakenthroughport3.
Oneendofthesensorhasa15mmmaleconnectorforaY-piececonnection,while
thepatientendconsistsofa22mmmaleconnectorfortheventilationmaskand
aconcentric15mmfemaleconnectorfortheendotrachealtube.Adisposable
double-lumentubeconductstheowsignalasapressuredifferencetothepressure
sensorinsidethemonitor.Thismeasurementmethodmeansthatthereisnogasow
inthedouble-lumentube,butonlypressurepick-up.Arespiratorygassamplingline
connectedtotheD-lite(+)/Pedi-lite(+)completesthemonitoringsystem.Gassampling
isimportanttoenablecompensationoftheeffectofdifferentgasviscositiesontidal
andminutevolumecalculations.
PatientSpirometryparameters
MeasuredPatientSpirometryparametersarethefollowing:
Inspiratoryandexpiratorytidalvolumes(TVinsp/exp)
Inspiratoryandexpiratoryminutevolumes(MVinsp/exp)
Airwaypressures
Peakpressure(Ppeak):maximumpressureduringonebreath
Plateaupressure(Pplat):pressureatthereversalpointoftheow
Meanpressure(Pmean):averagepressureduringonebreath
Real-timepressurewaveform(Paw)
Positiveendexpiratorypressure(PEEPtot)inORandPACUsoftwarepackagesonly:
thepressureinthelungsattheendofexpiration,measuredwhenexpiratoryphase
changestoinspiratoryow.
Extrinsicpositiveendexpiratorypressure(PEEPe)andintrinsicpositiveend
expiratorypressure(PEEPi)(notinORandPACUsoftwarepackages).
Compliance(Compl):calculatedforeachbreath,indicatesthepressuredifference
requiredtodeliveracertainvolumeofgasintothepatient.
2062971-001CARESCAPEModularMonitors329