Invasivepressures
DonotturnontheIABPalgorithmunlessaballoonpumpisinuse.
Iftwopressuremeasurementmodulesthatmaptothesamepressurechannel
areconnected,therstdetectedpressuremeasurementmoduleisassignedto
theindicatedpressurechannel.
Apressurechannelisactivatedwhenapressuretransducerinterfacecableis
connectedtothePDM(Masimo),PDM(Nellcor),Tram451,Tram451N5,Tram
451M,DualBP,BP/CO,BP,orBP/DualTemp,orE-COPmodule,orwhenapressure
transducerisconnectedtotheE-COPSv,E-PSMP,E-PP,E-P,orE-PTmodule.
Apressurechannelisdeactivatedwhenthepressuretransducerinterfacecableis
disconnectedfromthePDM(Masimo),PDM(Nellcor),Tram451,Tram451N5,Tram
451M,DualBP,BP/CO,BP,orBP/DualTemp,orE-COPmodule,orwhenthepressure
transducerisdisconnectedfromtheorE-COPSv,E-PSMP,E-PP,E-P,orE-PTmodule.
E-modules:Adeactivatedpressurechanneldoesnotreleasetheassignedpressure
channel.Toreleaseanassignedpressurechannel,removethemodulefromthe
system.
PCWP:Followyourcareunit’spolicyandproceduresforobtainingPCWP
measurements,includingballooninationduration.
SPVandPPV:Measurementisreliableonlywhenthepatientismechanically
ventilated.
Invasivepressuremeasurementsetup
Invasivepressureequipmenttopatientconnection
1.Modulewithinvasivepressuremeasurementcapability
2.Fluidbagwithpressureinfusor
3.Transducersetup
4.Invasivebloodpressureadaptercable;singleordualcable(optional)
Invasivepressuremodulekeys
Thereareinvasivepressurekeysonthefollowingmodules:
236CARESCAPEModularMonitors2062971-001