Alarms
2.SelecttheAlarmLimitstab.
3.SelectAutoLimits.
Ifyouneedtoundothesechangesandreturntothepreviousalarmlimitsettings,
selectUndoSettingsbeforeclosingthemenu.
Defaultautoalarmlimits
Parameter
HighlimitLowlimit
NIBP
S/D/M:NIBP*1.25+10S/D/M:NIBP*0.75-10
AllHR/PRparameters
(ECG,SpO
2
,UAC,Art,
ABP,Fem)
AllHR*1.25ofthecurrentHRvalue
(averagedoverlast10s)
AllHR*0.75ofthecurrentHR
value(averagedoverlast10s)
STgroup
Greatestvalueingroup:+1ifgroupis
enabled,otherwiselimitis+2
Smallestvalueingroup:-1ifgroup
enabled,otherwiselimitis-2
STindividualST+1ifenabled,otherwiselimitis+2ST-1ifenabled,otherwiselimitis-2
PVCPVC+10
Notapplicable
EtCO
2
EtCO
2
+5%EtCO
2
-5%
SpO
2
SpO
2
+5%SpO
2
-5%
Art,ABP,Fem,P1
Sys/Dia/Mean:
Value*1.25+10mmHg
Value*1.25+1.3kPa
Sys/Dia/Mean:
Value*0.75-10mmHg
Value*0.75-1.3kPa
FemV,CVP,PA,RAP,
RVP,LAP,ICP,CPP,
P2-8
Sys/Dia/Mean:
Value*1.25+5mmHg
Value*1.25+0.67kPa
Sys/Dia/Mean:
Value*0.75-5mmHg
Value*0.75-0.67kPa
UAC/UVCSys/Dia/Mean:
Value*1.25+5mmHg
Value*1.25+0.67kPa
Sys/Dia/Mean:
Value*0.75-5mmHg
Value*0.75-0.67kPa
SvO
2
SvO
2
+5%SvO
2
-5%
Temperature,TBlood
Tx+1°C
Tx+1.8°F
Tx-1°C
Tx-1.8°F
Tx-Ty(e.g.,T2-T1)
Tx-Ty+1°C
Tx-Ty+1.8°F
Tx-Ty-1°C
Tx-Ty-1.8°F
PpeakPpeak+10cmH
2
OPpeak-10cmH
2
O
PEEPtot
PEEPtot+5cmH
2
OPEEPtot-5cmH
2
O
PEEPe
PEEPe+5cmH
2
OPEEPe-5cmH
2
O
MVexp
MVexp+2MVexp-2
RR
RR*1.25+2RR*0.75-2
Returningthedefaultalarmlimits
1.SelectAlarmSetupfromthemonitor’smainmenu.
2.SelecttheAlarmLimitstab.
2062971-001CARESCAPEModularMonitors143