Impedancerespiration
WARNING
DEFIBRILLATORPRECAUTIONSPatientsignalinputs
labeledwiththeCFandBFsymbolswithpaddlesare
protectedagainstdamageresultingfromdebrillation
voltages.Toensureproperdebrillatorprotection,useonly
therecommendedcablesandleadwires.
WARNING
ELECTRODECONFIGURATIONImpedancerespiration
monitoringisnotreliablewhenECGelectrodesareplaced
anywherebutonthechest.
WARNING
ELECTRODESWheneverpatientdebrillationis
apossibility,usenon-polarizing(silver/silverchloride
construction)electrodesforECGmonitoring.Polarizing
electrodes(stainlesssteelorsilverconstructed)maycause
theelectrodestoretainaresidualchargeafterdebrillation.A
residualchargewillblockacquisitionoftheECGsignal.
WARNING
PDMandTRAMIftheCardiacArtifactAlarmisturned
off,apneaeventsmaynotbedetected.
Respirationcautions
CAUTION
Theimpedancerespirationmeasurementmaycauserate
changesinMinuteVentilationRateResponsivePacemakers.
Setthepacemakerrateresponsivemodeoff(PDMandTRAM
only)orturnofftheimpedancerespirationmeasurementon
themonitor.
Respirationmeasurementlimitations
PSM:Impedancerespirationisintendedforpatientsoverthreeyearsold.
Electricaldevices,suchaselectrosurgeryunitsandinfraredheaters,thatemit
electromagneticdisturbance,maycauseartifactsordisabletherespiration
measurementcompletely.
Movementartifacts,shivering,andinterferencefromtheheartmayinterferewith
therespirationmeasurement.
Respirationpointstonote
Connectonlyoneimpedancerespirationmoduletothemonitorsimultaneously.
Donotplaceelectrodesonobviouslayersoffat,ormajormuscles.
Makesuretheelectrodegelismoist.
Makesureelectrodeshavegoodskincontact.
Dependingontherespirationmoduleused,notallrespirationmeasurementsand
settingsareavailabletovieworchange.
SincerespirationmonitoringissocloselylinkedwithECGmonitoring,patient
preparationandelectrodeplacementareimportant.
Intermittentmechanicalventilation:Duringspontaneousbreathingtheventilator
mayattimessupportthepatient’sventilationwithanextrainspiration.Ifthese
ventilatorinspirationsaresubstantiallylargerthanthespontaneousbreaths,the
respirationcalculationmaymistakenlycountonlytheinspirationsandexpirations
producedbytheventilator.