Pulseoximetry
ChangingtheSpO
2
waveformscale
NOTE
PSMonly.
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2orSpO2(2)tab.
3.SelectthescalefromtheScalelist:
AUTO:ThescaleisautomaticallyselectedaccordingtotheIrMod%(infrared
modulationpercentage)receivedfromthemeasurementsource.
Otherscaleoptionsare2,5,10,20,or50.
SelectingtheSpO
2
hemodynamicsweepspeed
NOTE
Thissettingadjuststhewaveformspeedforallofthe
hemodynamicparameters.
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2orSpO2(2)tab.
3.SelectanumericvaluefromtheHemodynamicSweepSpeedlist.
Thesmallerthevalue,theslowerthesweepspeed.
SelectingtheSpO
2
astheprimaryheartratesource
TheprimaryheartratecanbecalculatedfromtheECGleads,SpO
2
measurement,
orinvasivepressurewaveform.
NOTE
Thissettingadjuststheprimaryheartratesourceforallofthe
hemodynamicparameters.
NOTE
HRAlarmsmustbeconguredasSingletoenabletheSpO
2
astheprimaryheartratesource.
NOTE
PrimarySpO
2
measurementonly.
SpO
2
canbethePrimaryHRSourceforallmodulesexcepttheTram-Racsingle
parametermodules.
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2tab.
3.SelecttheheartratesourcefromthePrimaryHRSourcelist.
ShowingtheSpO
2
pulserate
1.SelecttheSpO
2
parameterwindow.
2.SelecttheSpO2orSpO2(2)tab.
3.SelectShowPulseRate.
AdjustingtheSpO
2
pulsebeeptonevolume
Avariablepitchbeeptonerisesinpitchwithincreasingoxygensaturationorfallsin
pitchwithdecreasingoxygensaturation.
2062971-001CARESCAPEModularMonitors209