ECG
time,arrhythmiadetectionmaynotbeavailable.Ifthemonitorisnotabletorelearn
duetoalowamplitudeQRS,forexample,theArrhythmiapausedalarmistriggered.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelectRelearnQRS.
Automaticrelearningtakesplacewhen:
●Themeasurementmodechangesbetweenthe3–leadmodeandanyotherlead
mode.
●TheECG1Leadselectionischangedinthe3–leadmode.
●TheValeadselectionischangedinthe5–and6–leadmodes.
●TheECGcableisconnected(PSM,PDM).
●TheLeadAnalysissettingischangedfromMultileadtoSinglelead.
SettingtheprimaryHRsource
TheprimaryheartratecanbecalculatedfromtheECGleads,SpO
2
measurement,
orinvasivepressurewaveform.
NOTE
Thissettingadjuststheprimaryheartratesourceforallofthe
hemodynamicparameters.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelectaparameterfromthePrimaryHRSourcelist.Theselectionlistwillonly
showactivemeasurementsandAUTOorIntelliRate.Choicesare(UACwiththe
NICUsoftwarepackageonly):
●PDMwithsingleHR:IntelliRate,ECG,Art,ABP,Fem,UAC,Pleth.
●PDMwithmultipleHR:IntelliRate,ECG.
●E-modules,TRAM,telemetrytransmitterswithsingleHR:AUTO,ECG,Art,ABP,
Fem,UAC,Pleth.
●E-modules,TRAM,telemetrytransmitterswithmultipleHR:AUTO,ECG.
ShowingasecondHRvalueintheHRparameterwindow
YoucandisplayasecondheartratesourceintheHRparameterwindow.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelecttheShow2ndHRSourcecheckboxtodisplaythesecondHRsource.
●IftheprimaryHRsourceforE-modulesisECGorAUTO(ECG),thesecondaryHR
sourceisdisplayedinthisorder:Art,ABP,Fem,SpO2.
●IftheprimaryHRsourceforPDMisECGorIntelliRate(ECG),orAUTO(ECG)for
TRAM,thesecondaryHRsourcedisplayedinthisorder:UAC,Art,ABP,Fem,SpO2.
UACisavailableintheNICUsoftwarepackageonly.
●IftheprimaryHRsourceisanythingelsethanmentionedabove,thesecondaryHR
sourceisalwaysECG.
162CARESCAPEModularMonitors2062971-001