EasyManua.ls Logo

GE CARESCAPE - Selecting the HR alarm range

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECG
SettingHRalarmlimitsforasingleHRsource
1.SelecttheHRparameterwindow.
2.SelecttheAlarmstab.
3.Checkthattherequiredalarm,HR,PVC,orSVC,isturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SettingHR/PRalarmlimitsformultipleHRsources
1.SelecttheHRparameterwindow.
2.SelecttheHR/PRAlarmstab.
3.Checkthattherequiredalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedoutandtheAlarmOncheck
boxisnotselected.
4.SelecttheAlarmOncheckboxforthosealarmsyouwishtoset.
5.Adjustthealarmlimitswiththearrows.
SettingPVCalarmlimits
Availablewiththefullarrhythmialicenseonly.
1.SelecttheHRparameterwindow.
2.SelectthePVC/SVCAlarmsorPVCAlarms(TRAMortelemetrytransmitters)tab.If
theheartratesettingisSingle,selecttheAlarmstab.
3.CheckthatthePVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SettingSVCalarmlimits
Availablewiththefullarrhythmialicenseonly.Notavailablewithcombination
monitoringorTRAM.
1.SelecttheHRparameterwindow.
2.SelectthePVC/SVCAlarmstab.IftheheartratesettingisSingle,selectthe
Alarmstab.
3.CheckthattheSVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SelectingtheHRalarmrange
WiththeSingleheartratesettingonly.
164CARESCAPEModularMonitors2062971-001