ECG
SettingHRalarmlimitsforasingleHRsource
1.SelecttheHRparameterwindow.
2.SelecttheAlarmstab.
3.Checkthattherequiredalarm,HR,PVC,orSVC,isturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SettingHR/PRalarmlimitsformultipleHRsources
1.SelecttheHRparameterwindow.
2.SelecttheHR/PRAlarmstab.
3.Checkthattherequiredalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedoutandtheAlarmOncheck
boxisnotselected.
4.SelecttheAlarmOncheckboxforthosealarmsyouwishtoset.
5.Adjustthealarmlimitswiththearrows.
SettingPVCalarmlimits
Availablewiththefullarrhythmialicenseonly.
1.SelecttheHRparameterwindow.
2.SelectthePVC/SVCAlarmsorPVCAlarms(TRAMortelemetrytransmitters)tab.If
theheartratesettingisSingle,selecttheAlarmstab.
3.CheckthatthePVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SettingSVCalarmlimits
Availablewiththefullarrhythmialicenseonly.Notavailablewithcombination
monitoringorTRAM.
1.SelecttheHRparameterwindow.
2.SelectthePVC/SVCAlarmstab.IftheheartratesettingisSingle,selectthe
Alarmstab.
3.CheckthattheSVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntosetthealarms.
5.Adjustthealarmlimitswiththearrows.
SelectingtheHRalarmrange
WiththeSingleheartratesettingonly.
164CARESCAPEModularMonitors2062971-001