EasyManua.ls Logo

GE CARESCAPE - How to interpret the NMT values

GE CARESCAPE
548 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Neuromusculartransmission
ElectroSensorandelectromyography(EMG)
MeasurementwithElectroSensorisbasedonelectromyography.Threeelectrodesare
placedonthemuscle,andtheinnervatingnerveisstimulated.Therecordedelectrical
activityofthemuscleisinverselyproportionaltothedegreeofblock.
Stimulationmodes
Trainoffour,TOF:Recommendedformostcases.Itisalsothedefaultsetting.
Doubleburststimulation,DBS:UsefulwhenusingtheMechanoSensor.Itenables
bettervisualobservationofthefadinginresponses.
Posttetaniccount,PTC:Usedforestimatingtherelaxationlevelwithtetanic
stimulation.
Singletwitch,ST:Singletwitchmodeispracticalwhenusingdepolarizingrelaxants:
inthesecases,TOF%doesnotgiveanyadditionalinformationaboutthepatient
status.
HowtointerprettheNMTvalues
Whenneuromuscularblockdeepens,differentstimulationmodesmaybeneededto
assesstherelaxationstatus.Thefollowingtabledescribesthedepthofrelaxation.
100TOF%20
4
Count010PTC0
Light
RELAXATIONMETER
Deep
Trainoffour(TOF)mode
InTOFstimulusmode,fourstimulationpulsesaregeneratedat0.5secondintervals.
Theresponseismeasuredaftereachstimulusandtheratioofthefourthtotherst
responseoftheTOFsequenceiscalculated,resultinginTOF%.
WiththeElectroSensor,T1%isdisplayed.Ifthereferenceissuccessfullyfound,ascale
isalsoincluded.Scalemarkersrepresent0%,25%,50%,75%and100%reference
values.Whennoreferenceisavailable,noT1%valueisdisplayedandthebarsare
notscaled.
Whenrelaxationdeepens,theTOF%declinesuntilthefourthresponsedisappearsand
noTOF%isavailable.Thedegreeofneuromuscularblockisthenestimatedfromthe
numberofresponses,thecount,whichrepresentsthenumberofresponsesdetected
tothefourstimuli.Thefewertheresponses,thedeepertherelaxation.
Numberofresponses,countNeuromuscularblockMusclepower(%ofcontrol)
195%5%
290%10%
3
85%15%
4
75%25%
Doubleburststimulation(DBS)mode
DBSconsistsoftwoseparatebursts.Eachburstconsistsofthreeconsecutivepulses
ata50Hzfrequency.Theresponseratioofthesecondtotherstburstiscalculated,
resultinginDBS%(equivalenttoTOF%).