ECG
ShowingSTintheHRparameterwindow
ThisoptionisavailablewiththeMulti-leadSTAnalysislicenseonly.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelecttheShowSTcheckboxtodisplaySTintheHRparameterwindow.
ShowingPVCintheHRparameterwindow
ThisoptionisenabledwiththefullarrhythmialicenseandtheFulldetectionlevelonly.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelecttheShowPVCcheckboxtodisplayPVCintheHRparameterwindow.
ShowingQTintheHRparameterwindow
ThisoptionisavailablewiththeMulti-leadQT/QTcAnalysislicenseonly.
Thisselectionisnotavailablewithcombinationmonitoring.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelecttheShowQTcheckboxtodisplayQTintheHRparameterwindow.
DisplayingtheECGgrid
YoucanhaveareferencegridintheECG1,ECG2,andECG3waveformareas.Thegrid
pointswillbeat200mshorizontallyand0.5mVvertically.
1.SelecttheHRparameterwindow.
2.SelecttheAdvancedtab.
3.SelecttheECGGridcheckboxtodisplaythegrid.
ECGalarmlimits
TheHRAlarmscanbesettoSingleorMultiplethroughMonitorSetup>Default
Setup>CareUnitSettings>Parameters>ECG.Thissettingispasswordprotected.
TheSingleheartratesettingallowsyoutosetonecommonHRlimitformultiple
sources(e.g.,ECG,SpO
2
,Art)andthePVCandSVCalarmlimitsforECGfromthe
Alarmstab.Withthissettingactivated,turningofftheSpO
2
HRalarmlimitsalsoturns
offtheprimaryHRalarmandadjustingtheSpO
2
HRlimitvaluesalsoadjuststhe
primaryHRlimitvalue.
TheMultipleheartratesettingallowsyoutosetaprimaryheartrate/pulseratesource
anduptosixindividualheartrate/pulseratealarmsandlimitsfromtheHR/PRAlarms
tab.ItalsoallowsyoutosetPVCandSVCalarmlimitsforECGfromthePVC/SVC
Alarmstab.(ThetabiscalledPVCAlarmswithTRAMortelemetrytransmitters).With
theMultipleheartratesettingactivated,turningofftheSpO
2
HRalarmlimitsdoesnot
turnofftheprimaryHRalarm.
2062971-001CARESCAPEModularMonitors163