Message
Location
ExplanationWhattodo
Invivopoorsignal
param.
Themeasurementcannotbe
performedbecausethesignalis
tooweak.
Checkconnectionsand
catheter.
Changeopticalmoduleand
recalibrate.
Iftheproblempersists,the
cathetercouldbefaulty;
replacethecatheter.
Insufcientsignal
param.
Themeasurementcannotbe
performedbecausethesignalis
tooweak.
Checkthepatientstatus.
Checkconnectionsand
catheter.
Iftheproblempersists,
contactauthorizedservice
personnel.
Intensityshift
param.
Themeasurementcannotbe
performedbecausethesignal
intensitychanged.
Checkthepatientstatus.
Checkconnectionsand
catheter.
Iftheproblempersists,
contactauthorizedservice
personnel.
LowIntensity
param.
Catheteroatedorupagainst
vesselwall.
Checktheopticalmoduleand
connections.
Checkthecatheter
placement.
Onlyauthorizedmedical
personnelshouldadjust
catheterplacement.
LowLight
param.
Catheteroatedorupagainst
vesselwall.
Checktheopticalmoduleand
connections.
Checkthecatheter
placement.
Repeatinvivocalibration.
NoLight
param.
Interfaceddeviceshouldbe
checked.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
Re-calibrateSvO2al.area,
param.
Thecalibrationisover24hours
old.
Performinvivocalibration.
StartSvO2
param.
Displayswhenthecatheterhas
beeninsertedintothepatient
afterin-vitrocalibration.
SelecttheStartSvO2option
fromtheSvO
2
menu.
SvO2cableoffal.areaThecableisdisconnectedfrom
themodule.
Re-connectthecabletothe
module.
SvO2faultycableal.area,
param.
Theopticalmodulehasfailed.Checktheopticalmoduleand
connections.
Replacetheopticalmodule.
2062971-001CARESCAPEModularMonitors487