Messages
Message
Location
ExplanationWhattodo
●Invivopoorsignal
●param.
Themeasurementcannotbe
performedbecausethesignalis
tooweak.
●Checkconnectionsand
catheter.
●Changeopticalmoduleand
recalibrate.
●Iftheproblempersists,the
cathetercouldbefaulty;
replacethecatheter.
●Insufcientsignal
●param.
Themeasurementcannotbe
performedbecausethesignalis
tooweak.
●Checkthepatientstatus.
●Checkconnectionsand
catheter.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Intensityshift
●param.
Themeasurementcannotbe
performedbecausethesignal
intensitychanged.
●Checkthepatientstatus.
●Checkconnectionsand
catheter.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●LowIntensity
●param.
Catheteroatedorupagainst
vesselwall.
●Checktheopticalmoduleand
connections.
●Checkthecatheter
placement.
●Onlyauthorizedmedical
personnelshouldadjust
catheterplacement.
●LowLight
●param.
Catheteroatedorupagainst
vesselwall.
●Checktheopticalmoduleand
connections.
●Checkthecatheter
placement.
●Repeatinvivocalibration.
●NoLight
●param.
Interfaceddeviceshouldbe
checked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Re-calibrateSvO2●al.area,
param.
Thecalibrationisover24hours
old.
●Performinvivocalibration.
●StartSvO2
●param.
Displayswhenthecatheterhas
beeninsertedintothepatient
afterin-vitrocalibration.
●SelecttheStartSvO2option
fromtheSvO
2
menu.
●SvO2cableoff●al.areaThecableisdisconnectedfrom
themodule.
●Re-connectthecabletothe
module.
●SvO2faultycable●al.area,
param.
Theopticalmodulehasfailed.●Checktheopticalmoduleand
connections.
●Replacetheopticalmodule.
2062971-001CARESCAPEModularMonitors487