ExplanationWhattodo
●Calibrating
●param.
Calibrationfactorsarebeing
calculatedandstoredtothe
module.
●Noactionrequired.
●Checkcath.position
●param.
Lowsignalquality.●Checktheconnectionsand
catheter.
●CheckExtDevice
●param.
Interfaceddeviceshouldbe
checked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Connecting
●param.
Connectiontotheinterfaced
deviceisbeingestablished.
●Noactionrequired.
●DampedIntensity
●param.
Theremaybeproblemswith
connections.
●Checkthecatheter.
●Checkthecableand
connections.
●Recalibrateinvivo.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Drawblood
●param.
Anadvisoryprompt.●Drawbloodasindicatedon
screen.
●Enterlabresults
●param.
Calibrationsequencerequires
labresults.
●Enterlaboratoryresultsas
indicatedonscreen.
●HighIntensity
●param.
Catheteroatedorupagainst
vesselwall.
●Checktheopticalmoduleand
connections.
●Checkthecatheter
placement.
●Onlyauthorizedmedical
personnelshouldadjust
catheterplacement.
●Invitrocalibrating
●param.
Invitrocalibrationisinprocess.●Waituntilthemessage
disappears.
●Invitrofailed
●param.
Unsuccessfulcalibration.●Checkconnectionsand
catheter.
●Changeopticalmoduleand
recalibrate.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●Invivocalibrating
●param.
Invivocalibrationisinprocess.●Waituntilthemessage
disappears.