Calibrating
param.
Calibrationfactorsarebeing
calculatedandstoredtothe
module.
Noactionrequired.
Checkcath.position
param.
Lowsignalquality.Checktheconnectionsand
catheter.
CheckExtDevice
param.
Interfaceddeviceshouldbe
checked.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
Connecting
param.
Connectiontotheinterfaced
deviceisbeingestablished.
Noactionrequired.
DampedIntensity
param.
Theremaybeproblemswith
connections.
Checkthecatheter.
Checkthecableand
connections.
Recalibrateinvivo.
Iftheproblempersists,
contactauthorizedservice
personnel.
Drawblood
param.
Anadvisoryprompt.Drawbloodasindicatedon
screen.
Enterlabresults
param.
Calibrationsequencerequires
labresults.
Enterlaboratoryresultsas
indicatedonscreen.
HighIntensity
param.
Catheteroatedorupagainst
vesselwall.
Checktheopticalmoduleand
connections.
Checkthecatheter
placement.
Onlyauthorizedmedical
personnelshouldadjust
catheterplacement.
Invitrocalibrating
param.
Invitrocalibrationisinprocess.Waituntilthemessage
disappears.
Invitrofailed
param.
Unsuccessfulcalibration.Checkconnectionsand
catheter.
Changeopticalmoduleand
recalibrate.
Iftheproblempersists,
contactauthorizedservice
personnel.
Invivocalibrating
param.
Invivocalibrationisinprocess.Waituntilthemessage
disappears.
486CARESCAPEModularMonitors2062971-001