Messages
Message
Location
ExplanationWhattodo
Noise
wavef.Arrhythmiaalarmcategoryhas
beensettoOffintheECG,and
theECGchannelshavenotbeen
availableforanalysisforthe
last20secondsortheinternal
HRcalculationhasnotbeen
updatedforthelast30seconds
duetoexcessivenoisewhich
compromisestheaccuracyof
detectingevents.
Removethesourceof
excessivenoiseifpossible.
NoisyECG
al.areaArrhythmiaalarmcategoryhas
beensettoOffthroughECG>
Arrhythmia>LethalAlarms,
andtheECGchannelshave
notbeenavailableforanalysis
forthelast20secondsorthe
internalHRcalculationhas
notbeenupdatedforthelast
30secondsduetoexcessive
noisewhichcompromisesthe
accuracyofdetectingevents.
Checkandremovesourcesof
excessivenoise.
Pause
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Pleasewait-
collectingwaveform
report
ThereisnotenoughECGsample
datatorunthealgorithm.
Waitforabout10seconds.
QThighal.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
QTchighal.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
RonT
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
RemoveoneECG
module
al.areaTherearetwoECGmodulesin
thesystem.
RemoveoneECGmodule.
SinglePVCal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
STAnthigh/STAnt
low
STInfhigh/STInflow
STLathigh/STLat
low
al.area,
param.
Measurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
STXXXhigh/STXXX
low
whereXXX=ECGlead
label
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
466CARESCAPEModularMonitors2062971-001