Messages
Message
Location
ExplanationWhattodo
●Calibrating
●param.
Calibrationisinprogress.●Noactionrequired.
●Calibrationfail
●param.
Unsuccessfulcalibration.●Authorizedservicepersonnel
shouldrepeatthecalibration
procedure.
●CCIhigh/CCIlow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●CCOhigh/CCOlow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●CheckExtDevice
●param.
Interfaceddeviceshouldbe
checked.
●Checktheinterfaceddevice.
●Iftheproblempersists,
contactauthorizedservice
personnel.
●C.O.Complete
●C.O.
menu
Successfulcardiacoutput
determinationhasbeen
performed,butthemodule
isnotyetreadyforanew
measurement.Alsowhenthe15
minutetimeoutforconrming
theC.O.measurementexpires
withtheCardiacOutput>
Measurementmenuopen.
●Waituntilthemessage
disappears.
●COmeasurement
removed
●al.areaAcquisitionmodulehasbeen
removed.
●Connectthemoduleif
youwanttorestartthe
measurement.
●C.O.outofrange
●C.O.
menu
Themeasurementresultsare
invalid.
●Performanewmeasurement.
●ConrmC.O.●al.area,
param.
Measurementdatahasnot
beenconrmedbeforetrying
toexittheCardiacOutput>
Measurementmenu.
Ifmorethan15minuteshave
passedsincethestartoftheC.O.
measurement,thismessage
resetsautomatically.
●SelectConrmC.O.
●Curveoverrange●C.O.
menu
Thermodilutionwaveformvalues
areoverthevalidrange.
●Waitfortheblood
temperaturetostabilize
andretrythemeasurement.
●Curveunderrange
●C.O.
menu
Thermodilutionwaveformvalues
areunderthevalidrange.
●Waitfortheblood
temperaturetostabilize
andretrythemeasurement.
●IdenticalC.O.
modules
●al.areaTherearetwoormoreidentical
C.O.modulesinthesystem.
Alsowhentherearemorethan
oneactiveC.O.sources.
●RemoveallbutoneC.O.
module.
2062971-001CARESCAPEModularMonitors483