Message
Location
ExplanationWhattodo
Calibrating
param.
Calibrationisinprogress.Noactionrequired.
Calibrationfail
param.
Unsuccessfulcalibration.Authorizedservicepersonnel
shouldrepeatthecalibration
procedure.
CCIhigh/CCIlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
CCOhigh/CCOlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
CheckExtDevice
param.
Interfaceddeviceshouldbe
checked.
Checktheinterfaceddevice.
Iftheproblempersists,
contactauthorizedservice
personnel.
C.O.Complete
C.O.
menu
Successfulcardiacoutput
determinationhasbeen
performed,butthemodule
isnotyetreadyforanew
measurement.Alsowhenthe15
minutetimeoutforconrming
theC.O.measurementexpires
withtheCardiacOutput>
Measurementmenuopen.
Waituntilthemessage
disappears.
COmeasurement
removed
al.areaAcquisitionmodulehasbeen
removed.
Connectthemoduleif
youwanttorestartthe
measurement.
C.O.outofrange
C.O.
menu
Themeasurementresultsare
invalid.
Performanewmeasurement.
ConrmC.O.al.area,
param.
Measurementdatahasnot
beenconrmedbeforetrying
toexittheCardiacOutput>
Measurementmenu.
Ifmorethan15minuteshave
passedsincethestartoftheC.O.
measurement,thismessage
resetsautomatically.
SelectConrmC.O.
CurveoverrangeC.O.
menu
Thermodilutionwaveformvalues
areoverthevalidrange.
Waitfortheblood
temperaturetostabilize
andretrythemeasurement.
Curveunderrange
C.O.
menu
Thermodilutionwaveformvalues
areunderthevalidrange.
Waitfortheblood
temperaturetostabilize
andretrythemeasurement.
IdenticalC.O.
modules
al.areaTherearetwoormoreidentical
C.O.modulesinthesystem.
Alsowhentherearemorethan
oneactiveC.O.sources.
RemoveallbutoneC.O.
module.
2062971-001CARESCAPEModularMonitors483