ECG
AlarmmessageArrhythmiaanalysisAlarmprioritydefaultArrhythmiadetection
criteria
CoupletFullAccordingtopriority
setting
TwoconsecutivePVCs
aredetectedbetween
normalbeats,N-V-V-N.
Thecouplinginterval
betweenthePVCsmust
belessthan600ms.
IrregularFull
PDMinNICUsoftware
packageonly,TRAMwith
allsoftwarepackages
Accordingtopriority
setting
Sixconsecutivenormal
RRintervalsvaryby100
msormore.
MissingbeatFull
E-moduleswithall
softwarepackages;
PDMwithothersoftware
packagesexceptNICU
Accordingtopriority
setting
ActualRRintervalmore
than1.8timesthe
averageRRinterval.
MultifocalPVCsFull
E-modules,PDM
Accordingtopriority
setting
Overthelast15beats
twoormorePVCswith
differentmorphologies
aredetected.
Pause
FullAccordingtopriority
setting
Couplinginterval
betweentwobeats
exceeds:
3secondswithTRAM
andcombination
monitoring
1to5seconds
(congurable)with
E-modulesandPDM
RonT
FullAccordingtopriority
setting
IsolatedPVCisdetected
within100msofthe
peakoftheT-waveofthe
patient’spredominant
normalbeat.
SinglePVCFull
TRAMorcombination
monitoringonly.
Accordingtopriority
setting
IsolatedPVCisdetected.
SVTachyFull
E-modulesandPDMonly
Accordingtopriority
setting
ArunofSVCsisdetected
witharunlengthofat
leastthesetSVTLength
andtheheartrateisat
leastthesetHRforSVT
/min.
2062971-001CARESCAPEModularMonitors179