Pulseoximetry
WARNING
Pulseratemeasurementisbasedontheopticaldetectionof
aperipheralowpulseandthereforemaynotdetectcertain
arrhythmias.Thepulseoximetershouldnotbeusedasa
replacementorsubstituteforECG-basedarrhythmiaanalysis.
WARNING
Cable/sensoraftercare:
Donotreusesensorsintendedforsinglepatientuse.
Donotsterilizesensorsorpatientcablesbyirradiation,
steam,orethyleneoxide.
Cleanthesurfaceoftheprobebeforeandaftereach
patientuse.
Allowsensorandcabletodrycompletelyaftercleaning.
Moistureanddirtontheconnectorcanaffectthe
measurementaccuracy.
Ifaprobeisdamagedinanyway,discontinueuse
immediately.
InaccurateSpO
2
datacanresultifasensorispastitsuseful
life.Therefore,re-evaluatethemeasurementperiodically
byperformingadditionalassessmentofthepatientand
equipment,includingconsiderationofuseofalternate
monitoringmethodssuchasdirectmeasurementof
arterialoxyhemoglobinsaturation(SaO
2
).
Adamagedsensormaycauseburnsduringelectrosurgery.
WARNING
Oximetryperformancemaybeimpairedwhenpatient
perfusionisloworsignalattenuationishigh.
WARNING
B850only:CABLECOMPATIBILITYTram451NandTram
851NmodulesrequireNellcorOxismartXLcablesandsensors.
Older(non-OxismartXL)cablesmustnotbepluggedintothe
SpO
2
connectoronthesemodules.Useofnon-OxismartXL
cablesmayresultinerroneousreadings.
WARNING
NEONATALThedisplayofinaccuratepulseoximetry(SpO
2
)
valueshasbeenlinkedtothepresenceofpoorsignalstrength
orartifactduetopatientmotionduringsignalanalysis.This
conditionismostlikelytobeencounteredwhenthemonitoris
usedonneonatesorinfants.Thesesameconditionsinadults
donotimpacttheSpO
2
valuestothesameextent.
Werecommendtheapplicationofthefollowingcriteriawhen
usingthepulseoximetryfunctiononneonatesandinfants:
Theperipheralpulserate(PPR)asdeterminedbytheSpO
2
functionmustbewithin10%oftheheartrate,and
TheSpO
2
signalstrengthshouldbeadequate.Thisis
indicatedbythedisplayoftwoorthreeasterisksorthe
absenceoftheLowsignalqualitymessage.
Proceduresordevicespreviouslyappliedinyourfacilityfor
SpO
2
monitoringshouldbeusedintheeventtheSpO
2
value
fromthemonitorcannotbevalidatedbytheabovecriteria.
202CARESCAPEModularMonitors2062971-001