Cardiacoutput
ProblemSolution
WhatiftheC.O.valueishigherthanexpected?Cardiacoutputmustbecomputedwithin20
seconds.Increasingthevolumeanddecreasing
thetemperaturewillgiveyouagreaterdifferential
change.
Increasethevolumeinjected.
Decreasethetemperatureoftheinjectate.
Whatifastablebaselinetemperaturecannotbe
detected?
Amessageindicatingafailureisdisplayed.
CheckthepatientandtheC.O.setup(boththe
settingsandthecables).
Checkforasignicantamountofrespiratory
variationandforrapidIVsolutioninfusion,either
ofwhichmayinuencethebaselinetemperature.
Itmaybenecessarytostoporslowthesolution
infusionduringC.O.measurement,however,use
cautionifthesolutionincludesdrugs/medication.
Checktheinjectatetemperature(IT).Thereshould
beaminimumtemperaturedifferenceof10°C(18°
F)betweenthepatientbloodtemperatureandthe
injectatesolutiontemperature.Cooltheinjectate
solutionifneededtoincreasethedifference.
Replacetheinjectatetemperaturecable.
Thepulmonaryarterycathetermaybedamaged.
Replaceit.
Whatifthecardiacoutputvaluesareinaccurate?Whenin-lineisbeingusedalongwithicedinjectate,
theinitialtemperaturedisplayedwillbetheroom
temperature.However,whenthesolutionisinjected,
thetemperaturedisplayedwilldecrease.
Technique.Itisimportanttounderstandthe
techniqueusedinperformingacardiacoutput
sinceitisamajorinuencingfactorinobtaining
accuratecardiacoutputvalues.
Ifroomtemperaturesolutionisused,besure
thebagisnotexposedtoasupplemental
heatsourceortouchingothersolutionsor
equipment.Thisisimportantsothesolution
temperaturewillbethesameastheroomair
temperaturesensedthroughthebathorin-line
probe.Anydifferenceintemperaturecouldgive
aninaccuratereading.
Wheninjecting,alwaysholdthesyringebythe
plungerandnotbythebarrel.Thetemperature
ofthesolutionincreasesataslowerrateifthe
barrelisnotheld,andthereforereducesthe
potentialforerrorinacardiacoutputvalue.
Itisrecommendedthatyouinjectrapidly
andsmoothlyintotheproximalportofthe
pulmonaryarterycatheter,within4to5
seconds.
2062971-001CARESCAPEModularMonitors271