Systemintroduction
Tram-Rachemodynamicparameters
Parameter
BP/DualTemp*
DualBP
BP
BP/C.O.
IP
1211
*Themonitorsupportstheinvasivebloodpressuremeasurementonly
Gasmodules
Thegasacquisitionmodulesprovidemeasurementsforgasparameters,anesthetic
agents,PatientSpirometry,gasexchange,andmetabolics.Theyarecompatiblewith
theCARESCAPEModularMonitorsasfollows:
E-sCO,E-sCOV,E-sCAiO,E-sCAiOV:MonitorsB850,B650,B450
E-miniC:MonitorsB850,B650,B450
E-CO,E-COV,E-COVX,E-CAiO,E-CAiOV,E-CAiOVX:MonitorsB850,B650
Tram-Rac:MonitorB850
E-modulegasparameters
Module
CO
2
N
2
OO
2
Anesthetic
agentsAgentID
E-miniCx*---
E-sCOx*x--
E-sCOVx*x--
E-COx*x--
E-COV
x*x--
E-COVXx*x--
E-sCAiOxxxxx
E-sCAiOVxxxxx
E-CAiOxxxxx
E-CAiOVxxxxx
E-CAiOVXxxxxx
*TheE-sCO,E-CO,E-sCOV,E-COV,andE-COVXmodulesautomaticallycompensateforN2Oinrealtime
althoughN
2
Ovaluesarenotdisplayedonscreen.TheE-miniCrequiresmanualselectionfromthemonitor
menutocompensateforN
2
O.
ModulePatientSpirometryGasexchange
E-miniC--
E-sCO--
E-sCOVx-
E-CO--
E-COVx-
E-COVXxx
E-sCAiO--
2062971-001CARESCAPEModularMonitors75