Command Reference
R&S
®
ZNB/ZNBT
1037User Manual 1173.9163.02 ─ 55
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
11
.
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC9:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC9:Y:RPOS 50
or:
DISP:WIND2:TRAC:Y:RPOS 50, 'CH4TR1'
Set the reference position to the center of the diagram area.
Manual operation:
See "Ref Pos" on page 298
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet <MagnitudeFactor>[,
<PhaseFactor>, <RealPart>, <ImaginaryPart>]
Modifies all points of the trace <WndTr> by means of an added and/or a multiplied
complex constant. The response values M of the trace are transformed according to:
Suffix:
<Wnd>
.
Number of an existing diagram area (defined by means of
DISPlay[:WINDow<Wnd>][:STATe] ON).
<WndTr> Existing trace number, assigned by means of DISPlay[:
WINDow<Wnd>]:TRACe<WndTr>:FEED.
Parameters:
<MagnitudeFactor> Multiplied magnitude factor
Range: -300 dB to + 300 dB
*RST: 0 dB
Default unit: dB
<PhaseFactor> Multiplied phase factor, optional for setting command but
returned by query
Range: -3.4*1038 deg to +3.4*1038 deg
*RST: 0 deg
Default unit: deg
<RealPart> Real and imaginary part of added complex constant, optional for
setting command but returned by query
Range: -3.4*1038 to +3.4*1038
*RST: 0
<ImaginaryPart>
SCPI Command Reference