[SENSe<Ch>:]CORRection:METas:CABLe:REConnection [<Ports>, <Ports>...]
While a METAS calibration is active for the related channel (see [SENSe<Ch>:
]CORRection:METas[:STATe] on page 1133), this command notifies the METAS
VNA Tools II about additional connector reconnections at the specified ports.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Ports> <integer>
First port number, where connectors were reconnected
ALL
All ports were reconnected
NONE
Clears the connector reconnection history in the METAS project
<Ports> Additional port numbers, where connectors were reconnected
Usage: Setting only
Options: R&S ZNB-K50
Manual operation: See "Add Conn. Reconnection/Add Cable Movement"
on page 509
[SENSe<Ch>:]CORRection:METas:CABLe:TYPE <CableNameFile>[, <Ports>,
<Ports>...]
Selects the METAS cable type that is used at the respective ports.
Use [SENSe:]CORRection:METas:CABLe:CATalog? to list the available cable
types.
Suffix:
<Ch>
.
Channel number
The related channel must be prepared for a METAS calibration
using [SENSe<Ch>:]CORRection:COLLect:METHod:
DEFine <CalName>, MTOSm | MUOSm | MOSM |
MTRL, ....
Setting parameters:
<CableNameFile> Name of a METAS cable model file (without *.cable exten-
sion) in the database of the local METAS VNA Tools II installa-
tion.
<Ports> First port number or ALL, if the same cable type is used for all
ports.
The latter is equivalent to omitting all <Ports> parameters.
<Ports> Additional port numbers
Usage: Setting only
Options: R&S ZNB-K50
SCPI Command Reference