Command Reference
R&S
®
ZNB/ZNBT
1234User Manual 1173.9163.02 ─ 55
<Seg> Sweep segment number. This suffix is ignored.
Parameters:
<Boolean> ON – use independent GC settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol.
OFF – reset the GC mode in all sweep segments to the GC
mode for unsegmented sweeps, defined via [SENSe<Ch>:
]POWer:GAINcontrol or [SENSe<Ch>:]POWer:
GAINcontrol:GLOBal. ON will not restore the previous val-
ues.
*RST: ON
Example: See [SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
Manual operation: See "Segmented AGC" on page 543
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol <Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number. This suffix is ignored; the setting con-
trols the whole segmented sweep.
Parameters:
<Boolean> ON – use independent power settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel].
OFF – reset the power in all sweep segments to the power for
unsegmented sweeps, defined via SOURce<Ch>:
POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]. ON
will not restore the previous values.
The parameter is automatically switched to ON when a segment
power is entered using [SENSe<Ch>:]SEGMent<Seg>:
POWer[:LEVel].
*RST: OFF
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
SEGM:POW:CONT OFF
Couple the powers in all segments and reset the power in seg-
ment no. 1 to the initial value.
Manual operation: See "Optional Columns" on page 403
SCPI Command Reference