Command Reference
R&S
®
ZNB/ZNBT
1283User Manual 1173.9163.02 ─ 55
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt> Calibrated port number
Parameters:
<Boolean> Enables (ON) or disables (OFF) the source power calibration for
port number <PhyPt>.
*RST: OFF
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation: See "Port Overview" on page 497
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient[:STATe] <Boolean>
Enables or disables the use of two-port transmission coefficients.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean> Enables (ON) or disables (OFF) the two-port transmission coeffi-
cients.
*RST: n/a
Example: See SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration
Manual operation: See "Test Setup" on page 493
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:CALibration <Boolean>
Selects the position of the additional two-port in the test setup.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean> ON – two port at power meter (during calibration)
OFF – two port at DUT (during measurement)
*RST: ON (notice that the correction is switched off after a
reset)
SCPI Command Reference